Evaluation of Corneal Deformation Analyzed with Scheimpflug Based Device in Healthy Eyes and Diseased Ones
نویسندگان
چکیده
This study was designed to evaluate the correlation between corneal biomechanical and morphological data in healthy eyes, eyes that underwent myopic photorefractive keratectomy (PRK), keratoconus affected eyes, and keratoconus affected eyes that underwent corneal collagen crosslinking (CCC). Complete clinical eye examination of all eyes was followed by tomographic (Pentacam, Oculus, Wetzlar, Germany) and biomechanical (Corvis ST, Oculus, Wetzlar, Germany) evaluation. Differences among Corvis ST (CST) parameters in the different groups have been performed. Linear regression between central corneal thickness (CCT), intraocular pressure (IOP), and anterior corneal curvature measured with Sim'K (KM), versus corneal deformation parameters measured with Corvis ST in the different groups, has been run using SPSS software version 18.0. We evaluated 64 healthy eyes of 64 patients with a mean refractive error of -0.65 ± 1.68 D (measured as spherical equivalent), 17 eyes of 17 patients that underwent myopic PRK for a mean refractive defect of -4.91 ± 2.05 D (measured as spherical equivalent), 16 eyes of 16 patients affected by keratconus (stage 2-3 of Amsler Classification), and 13 eyes of 13 patients affected by keratoconus that underwent CCC. Our data suggest that corneal curvature would have a greater influence on corneal deformation than CCT; in fact KM values are more strongly associated with more CST parameters both about corneal change in shape and both about the corneal ability to come back at original shape.
منابع مشابه
مقایسهی کراتومتری پنتاکم وکراتومتری IOL Master در محاسبهی لنز داخل چشمی افراد با قرنیهی طبیعی
Background and Objective: Proper method and machine for corneal evaluation is an important factor in many anterior segment interventions. This study was performed to compare the mean keratometry (K) readings obtained with an IOL Master and Pentacam for intraocular lens power calculation in normal subjects with no history of refractive surgery. Materials and Methods: Mean K values were obtained ...
متن کاملFactors Associated With Corneal Deformation Responses Measured With a Dynamic Scheimpflug Analyzer.
Purpose The purpose of this study was to clarify correlations between corneal deformation parameters measured with a dynamic Scheimpflug analyzer and baseline factors such as axial length, intraocular pressure (IOP), age, central corneal thickness (CCT), and corneal curvature. Methods Ninety-six eyes of 96 healthy subjects (mean 55.2 ± 16.1 years of age) were examined using a dynamic Scheimpf...
متن کاملQuantitative Analysis of Corneal Energy Dissipation and Corneal and Orbital Deformation in Response to an Air-Pulse in Healthy Eyes.
PURPOSE To examine and evaluate ocular biomechanical metrics and additionally derived corneal and orbital components using a noncontact Scheimpflug-based tonometer (CorVis ST) in a population of healthy eyes. METHODS A total of 152 eyes of 152 participants were examined by slit-lamp biomicroscopy, corneal tomography, and the CorVis ST (CST). This determined the distribution of outputs from th...
متن کاملRepeatability of aberrometric measurements in normal and keratoconus eyes using a new Scheimpflug-Placido topographer.
PURPOSE To evaluate the repeatability of the anterior and posterior corneal wavefront aberrations using the Sirius Scheimpflug-Placido topographer in normal eyes and keratoconus eyes. SETTING Bozok University Faculty of Medicine, Yozgat, Turkey. DESIGN Evaluation of diagnostic test. METHODS In eyes of healthy subjects and eyes of keratoconus patients, 3 repeated measurements were obtained...
متن کاملComparison of central corneal thickness with four noncontact devices: An agreement analysis of swept-source technology
PURPOSE The purpose of this study was to compare the central corneal thickness (CCT) measurements of four noncontact devices in healthy eyes. MATERIALS AND METHODS In a sample of 45 healthy controls, CCT was measured using an optical biometer (IOLMaster 700) based on swept-source optical coherence tomography (SS-OCT), high-resolution rotating Scheimpflug camera system (Pentacam HR), spectral-...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 2014 شماره
صفحات -