Clinical Application of Gamma Knife Dose Verification Method in Multiple Brain Tumors : Modified Variable Ellipsoid Modeling Technique
نویسندگان
چکیده
OBJECTIVE The Leksell Gamma Knife® (LGK) is based on a single-fraction high dose treatment strategy. Therefore, independent verification of the Leksell GammaPlan® (LGP) is important for ensuring patient safety and minimizing the risk of treatment errors. Although several verification techniques have been previously developed and reported, no method has ever been tested statistically on multiple LGK target treatments. The purpose of this study was to perform and to evaluate the accuracy of a verification method (modified variable ellipsoid modeling technique, MVEMT) for multiple target treatments. METHODS A total of 500 locations in 10 consecutive patients with multiple brain tumor targets were included in this study. We compared the data from an LGP planning system and MVEMT in terms of dose at random points, maximal dose points, and target volumes. All data was analyzed by t-test and the Bland-Altman plot, which are statistical methods used to compare two different measurement techniques. RESULTS No statistical difference in dose at the 500 random points was observed between LGP and MVEMT. Differences in maximal dose ranged from -2.4% to 6.1%. An average distance of 1.6 mm between the maximal dose points was observed when comparing the two methods. CONCLUSION Statistical analyses demonstrated that MVEMT was in excellent agreement with LGP when planning for radiosurgery involving multiple target treatments. MVEMT is a useful, independent tool for planning multiple target treatment that provides statistically identical data to that produced by LGP. Findings from the present study indicate that MVEMT can be used as a reference dose verification system for multiple tumors.
منابع مشابه
Gamma Knife Simulation Using the MCNP4C Code and the Zubal Phantom and Comparison with Experimental Data
Introduction: Gamma Knife is an instrument specially designed for treating brain disorders. In Gamma Knife, there are 201 narrow beams of cobalt-60 sources that intersect at an isocenter point to treat brain tumors. The tumor is placed at the isocenter and is treated by the emitted gamma rays. Therefore, there is a high dose at this point and a low dose is delivered to the normal tissue surroun...
متن کاملEvaluation of Organ Specific peripheral dose for Gamma knife 4C based on Monte Carlo
Introduction: Stereotactic Gamma Knife radiosurgery has been widely used for treating brain tumors. The scattered radiation outside of treatment field (peripheral dose) can induce the secondary cancer to specific organ. This paper investigated the absorbed dose to eyes, thyroid, heart, lung, breast and colon using a Monte Carlo technique for Mird phantom. We also study the ef...
متن کاملNursing perspective of gamma knife of treatment
Gamma knife is a new noninvasive method for treatment of patient with brain tumors. ‎Another name of this procedure is steriotatic radiosurgery.‎‏ ‏This computerized advanced ‎technique was implicated by a team including neurosurgeon, neuroradialogist, oncologist, and ‎a nurse. This method is used for treatment of certain brain tumors and lesions. The benefits are ‎l...
متن کاملMonte Carlo Evaluation of Gamma Knife Dose Profile in Real Brain Phantom
Introduction The Gamma Knife system is designed solely for non-invasive treatment of brain disorders, and it benefits from stereotactic surgical techniques. Dose calculations required in the system are performed by GammaPlan code; in this code, brain tissue is considered uniform. In the present study, we evaluated the effect of Gamma Knife system on the obtained dose through simulating a real ...
متن کاملComparison dose distributions from gamma knife unit 4C with CT data and non-CT data options of beamnrc code
Todays gamma knife radiosurgery is used widely for treatment of very small brain tumors. In order to investigate accuracy of dosimetry and treatment planning calculations, using Monte Carlo simulation with dedicated code named as beamnrc including non-CT data and CT data options is necessary. The aim of this study is choosing the best options in order to have an accurate tools based on their ad...
متن کامل