IMRT verification with a camera-based electronic portal imaging system.

نویسندگان

  • M Partridge
  • J R Symonds-Tayler
  • P M Evans
چکیده

An evaluation of the capabilities of a commercially available camera-based electronic portal imaging system for intensity-modulated radiotherapy verification is presented. Two modifications to the system are demonstrated which use a novel method to tag each image acquired with the delivered dose measured by the linac monitor chamber and reduce optical cross-talk in the imager. A detailed performance assessment is presented, including measurements of the optical decay characteristics of the system. The overall geometric accuracy of the system is determined to be +/-2.0 mm, with a dosimetric accuracy of +/-1.25 MU. Finally a clinical breast IMRT treatment, delivered by dynamic multileaf collimation, is successfully verified both by tracking the position of each leaf during beam delivery and recording the integrated intensity observed over the entire beam.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Camera based EPID dosimetric verification of radiation treatments

Introduction: Tumor control and treatment quality depend on accuracy of delivered dose. Various methods have been studied for treatment dosimetry verification. Among these methods, the use of electronic portal imaging device (EPID) is highly considered. Many efforts are underway to improve the results of this context.   Materials and Methods: In this study, ...

متن کامل

EPID in vivo Dosimetry

Introduction: The most modern radiotherapy devices are equipped with an Electronic Portal Imaging Device (EPID) system which is located on opposite side of the machine’s head. EPID system is often used to setting up the position verification during or between radiotherapy sessions. Material and Methods: Various types of dosimeters have been used to setting up ...

متن کامل

Pretreatment quality control of single isocenter half- beam treatment planning technique using an amorphous silicon electronic portal-imaging device (EPID)

Introduction: Electronic portal imaging devices (EPIDs) are fundamentally used for instantaneous verification of the patient set‐up, block shape, and leaf positions during radiation therapy. In radiotherapy, situations arise in which an inclined PTV must be treated mutually with adjacent nodal regions. This methodology is most widely used for matching tangential/lateral breas...

متن کامل

Evaluation of a fast method of EPID-based dosimetry for IMRT and Comparison with 3D EPID-based dosimetry system using conventional two- and three-dimensional detectors for VMAT

Introduction: Electronic portal imaging devices (EPIDs) could potentially be useful for intensity-modulated radiation therapy (IMRT) and VMAT QA. The data density, high resolution, large active area, and efficiency of the MV EPID make it an attractive option. However, EPIDs were designed as imaging devices, not dosimeters, and as a result they do not inherently measure dose in ...

متن کامل

Patient Specific IMRT Pre-Treatment Verification Using A Commercial Portal Dosimetry Software

Introduction: Electronic portal imaging devices (EPIDs) are extensively used for dosimetric information of pre-treatment field verification and in-vivo dosimetry for Intensity Modulated Radiotherapy (IMRT). In the present work, we have implemented the newly developed portal dosimetry software using independent dose prediction algorithm EPIDose and evaluated this new tool for the pretreatment IM...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Physics in medicine and biology

دوره 45 12  شماره 

صفحات  -

تاریخ انتشار 2000