Reproducibility of Carotid Artery Doppler Frequency
نویسندگان
چکیده
Doppler sonography of 29 extracranial carotid arteries was performed twice within three days or less. Angiography revealed stenoses with from 10 to 95% diameter reduction in 25 internal carotid arteries whereas four vessels were found to be normal. The systolic peak frequency of the internal carotid artery read from the Doppler spectrum (n = 29) could be reproduced very well. This was shown by a linear regression nearest the line of identical values with a coefficient of correlation r = 0.97 (p < 0.001). Other values derived from the spectral analysis of the Doppler shift signal were not so well reproduced including the peak frequency ratio (systolic peak frequency of the internal carotid artery/systolic peak of the common carotid artery) (n = 22; r = 0.81; p < 0.001). The mean frequencies read from the zero-crossing detector recordings (n = 20) could not be reproduced as demonstrated by a linear regression far away from the line of identical measurements with a coefficient of correlation r = 0.43 (p < 0.05).
منابع مشابه
بررسی 291 مورد سونوگرافی داپلر کاروتید در بیماران کاندید CABG
Introduction: Doppler ultrasonography (DUS) is the most preferable method for screening of carotid artery disease in patients undergoing CABG. The purpose of this study was to investigate the frequency of carotid artery disease and determine the relation between known risk factors of atherosclerosis and rate of carotid artery narrowing in order to identify high-risk groups among patients schedu...
متن کاملReproducibility of carotid artery Doppler frequency measurements.
Doppler sonography of 29 extracranial carotid arteries was performed twice within three days or less. Angiography revealed stenoses with from 10 to 95% diameter reduction in 25 internal carotid arteries whereas four vessels were found to be normal. The systolic peak frequency of the internal carotid artery read from the Doppler spectrum (n = 29) could be reproduced very well. This was shown by ...
متن کاملCandida patients in a Jorjani center in Shahid Mohammadi Hospital of carotid Doppler sonnograghy findings in coronary artery Bypas grafting Bandar Abbas
Introduction: Nowadays caotid Dopplers sonography is gold standard for carotid artery screening in coronary aetery bypass grafting (CABG) candidates. This study was carried out in an angioplasty center in southern seaport in Iran to assesss theprevalence of artery stenosis and correlation of the atherosclerosis risk factor with the rate of artery stensis, and nence to define the high risk group...
متن کاملبررسی یافته های سونوگرافی داپلر عروق کاروتید اکستراکرانیال بیماران مبتلا به سکته مغزی (Anterior circulation) بر اساس محدوده گرفتاری در سی تی اسکن
Background and purpose : Ârterial embolism is one of the most prevalent cause of transient ischemic attack (TÏÂ) and other types of brain ischemias and about 5o% of TÏÂ patiens have vascular stenosis. Therefore the presence of mild atherosclerotic plaques in carotid arteries, as a source of embdlism in causing TÏÂ and other brain strokes are of a great importance. Ïn present study the ...
متن کاملارزیابی و بررسی نتایج حاصل از داپلر فراصوتی در مقایسه با آنژیوگرافی تفریقی دیجیتال در تشخیص تنگ شدگیهای شریان کاروتید داخلی
Introduction: Digital Subtraction Angiography (DSA) is the standard method for detection of carotid artery stenosis. This technique has high radiation dose. With the progress of imaging techniques, noninvasive or minimally invasive methods like the CT angiography, MR angiography and ultrasonography with Doppler or Duplex mode were applicable. This study aimed to evaluate and assesse the results...
متن کامل