Rationale and design of the HepZero study: a prospective, multicenter, international, open, randomized, controlled clinical study with parallel groups comparing heparin-free dialysis with heparin-coated dialysis membrane (Evodial) versus standard care: study protocol for a randomized controlled trial
نویسندگان
چکیده
BACKGROUND Anticoagulation for chronic dialysis patients with contraindications to heparin administration is challenging. Current guidelines state that in patients with increased bleeding risks, strategies that can induce systemic anticoagulation should be avoided. Heparin-free dialysis using intermittent saline flushes is widely adopted as the method of choice for patients at risk of bleeding, although on-line blood predilution may also be used. A new dialyzer, Evodial (Gambro, Lund, Sweden), is grafted with unfractionated heparin during the manufacturing process and may allow safe and efficient heparin-free hemodialysis sessions. In the present trial, Evodial was compared to standard care with either saline flushes or blood predilution. METHODS The HepZero study is the first international (seven countries), multicenter (10 centers), randomized, controlled, open-label, non-inferiority (and if applicable subsequently, superiority) trial with two parallel groups, comprising 252 end-stage renal disease patients treated by maintenance hemodialysis for at least 3 months and requiring heparin-free dialysis treatments. Patients will be treated during a maximum of three heparin-free dialysis treatments with either saline flushes or blood predilution (control group), or Evodial. The first heparin-free dialysis treatment will be considered successful when there is: no complete occlusion of air traps or dialyzer rendering dialysis impossible; no additional saline flushes to prevent clotting; no change of dialyzer or blood lines because of clotting; and no premature termination (early rinse-back) because of clotting.The primary objectives of the study are to determine the effectiveness of the Evodial dialyzer, compared with standard care in terms of successful treatments during the first heparin-free dialysis. If the non-inferiority of Evodial is demonstrated then the superiority of Evodial over standard care will be tested. The HepZero study results may have major clinical implications for patient care. TRIAL REGISTRATION ClinicalTrials.gov NCT01318486.
منابع مشابه
Results of the HepZero study comparing heparin-grafted membrane and standard care show that heparin-grafted dialyzer is safe and easy to use for heparin-free dialysis.
Heparin is used to prevent clotting during hemodialysis, but heparin-free hemodialysis is sometimes needed to decrease the risk of bleeding. The HepZero study is a randomized, multicenter international controlled open-label trial comparing no-heparin hemodialysis strategies designed to assess non-inferiority of a heparin grafted dialyzer (NCT01318486). A total of 251 maintenance hemodialysis pa...
متن کاملComparison of percutaneous versus open surgical techniques for placement of peritoneal dialysis catheter in children: A randomized clinical trial
Background: This research compares the outcomes of percutaneous technique and open surgical peritoneal dialysis catheter placement in children. Methods: In this randomized controlled trial, between 2010 and 2011,a total of 35 pediatric uremic patients were enrolled and randomized into two study groups. Follow up data included duration of operation (minute), duration of hospitalization (days...
متن کاملRationale and design of the HEALTHY-CATH trial: A randomised controlled trial of Heparin versus EthAnol Lock THerapY for the prevention of Catheter Associated infecTion in Haemodialysis patients
BACKGROUND Catheter-related bacteraemias (CRBs) contribute significantly to morbidity, mortality and health care costs in dialysis populations. Despite international guidelines recommending avoidance of catheters for haemodialysis access, hospital admissions for CRBs have doubled in the last decade. The primary aim of the study is to determine whether weekly instillation of 70% ethanol prevents...
متن کاملHemodialysis without Systemic Anticoagulation: A Prospective Randomized Trial to Evaluate 3 Strategies in Patients at Risk of Bleeding
OBJECTIVE In this clinical trial, we aimed to compare three means of performing chronic hemodialysis in patients with contra-indication to systemic heparinization. METHODS This open-label monocentric randomized « n-of-one » trial, conducted in a single tertiary care center, recruited chronic hemodialysis patients with a contra-indication to systemic heparinization for at least 3 consecutive s...
متن کاملThe clinical evaluation of low-dose heparin in haemodialysis: a prospective study using the heparin-coated AN69 ST membrane.
BACKGROUND The AN69 ST haemodialysis membrane, a new membrane resulting from coating polyethyleneimine upon the polyacrylonitrile surface, binds heparin. In patients at risk of bleeding, a pilot study has demonstrated the efficient anticoagulant effect of this heparin-coated membrane. Study design. In chronic haemodialyzed patients, we evaluated whether this anticoagulant effect can be validate...