Validating dose rate calibration of radiotherapy photon beams through IAEA/WHO postal audit dosimetry service.

نویسندگان

  • Abdul Qadir Jangda
  • Sherali Hussein
چکیده

In external beam radiation therapy (EBRT), the quality assurance (QA) of the radiation beam is crucial to the accurate delivery of the prescribed dose to the patient. One of the dosimetric parameters that require monitoring is the beam output, specified as the dose rate on the central axis under reference conditions. The aim of this project was to validate dose rate calibration of megavoltage photon beams using the International Atomic Energy Agency (IAEA)/World Health Organisation (WHO) postal audit dosimetry service. Three photon beams were audited: a 6 MV beam from the low-energy linac and 6 and 18 MV beams from a dual high-energy linac. The agreement between our stated doses and the IAEA results was within 1% for the two 6 MV beams and within 2% for the 18 MV beam. The IAEA/WHO postal audit dosimetry service provides an independent verification of dose rate calibration protocol by an international facility.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The IAEA/WHO TLD postal programme for radiotherapy hospitals.

BACKGROUND AND PURPOSE Since 1969 the International Atomic Energy Agency (IAEA), together with the World Health Organization (WHO), has performed postal TLD audits to verify the calibration of radiotherapy beams in developing countries. MATERIALS AND METHODS A number of changes have recently been implemented to improve the efficiency of the IAEA/WHO TLD programme. The IAEA has increased the n...

متن کامل

QA networks for radiotherapy dosimetry

Several national and international organizations have developed various types and levels of external audits for radiotherapy dosimetry, either based on on-site review visits or using mailed dosimetry systems. Three major TLD (thermoluminescent dosimetry) networks make available postal dose audits to a large number of radiotherapy centres on a regular basis. These are the IAEA/WHO TLD postal dos...

متن کامل

Postal dose audits for radiotherapy centers in Latin America and the Caribbean: trends in 1969-2003.

Since 1969 the International Atomic Energy Agency and the World Health Organization (along with the Pan American Health Organization, working with countries in Latin America and the Caribbean) have operated postal dosimetry audits based on thermoluminescent dosimetry (TLD) for radiotherapy centers. The purpose of these audits is to provide an independent dosimetry check of radiation beams used ...

متن کامل

Air Kerma Based Dosimetry versus Absorbed Dose to Water Based Dosimetry for High-energy Photon

Introduction: In the last 5 years the American Association of Physicists in Medicine Task Group 51 (AAPM TG-51) and the International Atomic Energy Agency (IAEA) published a new high-energy photon and electron dosimetry protocol. These protocols are based on the use of an ion chamber having an absorbed-dose to water calibration factor. These are different from the previous NCS report-2 and IAEA...

متن کامل

Dosimetric inter-institutional comparison in European radiotherapy centres: Results of IAEA supported treatment planning system audit.

BACKGROUND AND PURPOSE One of the newer audit modalities operated by the International Atomic Energy Agency (IAEA) involves audits of treatment planning systems (TPS) in radiotherapy. The main focus of the audit is the dosimetry verification of the delivery of a radiation treatment plan for three-dimensional (3D) conformal radiotherapy using high energy photon beams. The audit has been carried ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • JPMA. The Journal of the Pakistan Medical Association

دوره 62 5  شماره 

صفحات  -

تاریخ انتشار 2012