CRT-D implantation: one procedure, two problems

نویسندگان

  • Anna M. Engel
  • Jakub Machejek
  • Andrzej Ząbek
  • Artur Kozanecki
  • Jacek Lelakowski
  • Barbara Małecka
چکیده

A 70-year-old patient with chronic ischemic cardiomy-opathy (NYHA III, left ventricle ejection fraction (EFLV) = 15%, left bundle branch block (LBBB), percutaneous coronary intervention (PCI) left anterior descending artery (LAD) in 2006 and right coronary artery (RCA) in 2007), after an episode of sustained VT, terminated pharmacologically , was qualified for a cardiac resynchronization therapy (CRT-D) procedure (class IA). Reduction of heart failure was impossible despite optimal pharmacotherapy. The defibrillation lead was implanted in the right ventricle apex through the left cephalic venotomy, and the atrial lead was implanted in the right atrium appendage by a subclavian vein approach. Optimal atrial and ventricular pacing parameters were obtained. The left ventricular (LV) lead was implanted through left subclavian venepuncture with difficulties of coronary sinus (CS) cannulation. The cannulation of the vein running on the front wall of the heart – the marginal vein (M. von Ludinghausen type III of CS anatomy – Figure 1) – was enabled by using an LV introducing set with small curvature (Medtronic Attain 6216A-MB2). Administered contrast visualized single tributaries and the CS trunk. Cannulation of the CS was performed successfully based on the image of performed venography, followed by implantation of the LV lead in the wide lateral vein. Excellent pacing parameters were obtained in the distal location , although associated with diaphragm stimulation. More proximal reposition helped to gain acceptable LV stimulation parameters without any side effects. While the introducing sheath was removed, LV lead was dislocated with loss of effective stimulation. Instead of the previous standard Medtronic guide-wire, a hemodynamic guidewire (Abbott Hi-Torque Balance Heavyweight – Abbott-GW) was inserted into the lead, with the help of which an ineffective attempt of lead reposition was performed, using the over-the-wire technique. Due to earlier difficulties with CS cannulation we decided to introduce a new sheath using Abbott-GW, the ending of which was already in the CS. To do this part of the procedure it was necessary to remove the lead and leave only the Abbott-GW. The sheath was inserted, the Abbott-GW removed, and the LV lead was implanted again in the lateral vein with optimal electrical parameters. The sheath and the standard guidewire were removed without lead reposition. After the procedure was accomplished, reduction of the Abbott-GW's length (25 cm ending with fragile structure) was noticed. In a chest X-ray the fragment of the guidewire was localized between the superior vena cava (SVC) and pulmonary trunk …

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Success rate of transvenous left ventricular lead implantation for cardiac resynchronisation therapy - recent experience of a single centre.

BACKGROUND Implantation of a left ventricular (LV) lead for cardiac resynchronisation therapy (CRT) may be challenging. Wider use of various implantation techniques increases the success rate of CRT. AIM Short-term analysis of the success rate of transvenous LV lead implantation for CRT. METHODS All CRT procedures performed in 2009 with first-time LV lead implantation attempt were analysed ...

متن کامل

Anatomical factors involved in difficult cardiac resynchronization therapy procedure: a non-invasive study using dual-source 64-multi-slice computed tomography.

AIMS In cardiac resynchronization therapy (CRT) procedure, left ventricular (LV) lead implantation is time consuming. In this clinical setting, no study has investigated the impact of right atrium anatomical parameters on both CRT implantation procedure duration and X-ray exposure. Additionally, only few studies have examined the coronary sinus (CS) using dual-source 64-multi-slice computed tom...

متن کامل

Clinical decision tool for CRT-P vs. CRT-D implantation: Findings from PROSE-ICD

BACKGROUND Cardiac resynchronization therapy (CRT) devices reduce mortality through pacing-induced cardiac resynchronization and implantable cardioverter defibrillator (ICD) therapy for ventricular arrhythmias (VAs). Whether certain factors can predict if patients will benefit more from implantation of CRT pacemakers (CRT-P) or CRT defibrillators (CRT-D) remains unclear. METHODS AND RESULTS W...

متن کامل

Use of a quadripolar left ventricular lead to achieve successful implantation in patients with previous failed attempts at cardiac resynchronization therapy

AIMS Problems with implanting a left ventricular (LV) lead during cardiac resynchronization therapy (CRT) procedures are not uncommon and may occur for a variety of reasons including phrenic nerve stimulation (PNS) and high capture thresholds. We aimed to perform successful CRT in patients with previous LV lead problems using the multiple pacing configurations available with the St Jude Quartet...

متن کامل

Long-term retention of cardiac resynchronization therapy.

OBJECTIVES The purpose of this study was to determine the frequency and causes of intermittent and permanent loss of cardiac resynchronization therapy (CRT) in patients who have undergone the successful implantation of a transvenous defibrillator that delivers CRT (CRT-D). BACKGROUND The causes of loss of CRT have not been described. METHODS The records of 512 patients who underwent an atte...

متن کامل

Longevity of implantable cardioverter-defibrillators for cardiac resynchronization therapy in current clinical practice: an analysis according to influencing factors, device generation, and manufacturer

AIMS Device replacement at the time of battery depletion of implantable cardioverter-defibrillators (ICDs) may carry a considerable risk of complications and engenders costs for healthcare systems. Therefore, ICD device longevity is extremely important both from a clinical and economic standpoint. Cardiac resynchronization therapy defibrillators (CRT-D) battery longevity is shorter than ICDs. W...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 12  شماره 

صفحات  -

تاریخ انتشار 2016