Patient QA systems for rotational radiation therapy: a comparative experimental study with intentional errors.
نویسندگان
چکیده
PURPOSE The purpose of the present study was to investigate the ability of commercial patient quality assurance (QA) systems to detect linear accelerator-related errors. METHODS Four measuring systems (Delta(4®), OCTAVIUS(®), COMPASS, and Epiqa™) designed for patient specific quality assurance for rotational radiation therapy were compared by measuring four clinical rotational intensity modulated radiation therapy plans as well as plans with introduced intentional errors. The intentional errors included increasing the number of monitor units, widening of the MLC banks, and rotation of the collimator. The measurements were analyzed using the inherent gamma evaluation with 2% and 2 mm criteria and 3% and 3 mm criteria. When applicable, the plans with intentional errors were compared with the original plans both by 3D gamma evaluation and by inspecting the dose volume histograms produced by the systems. RESULTS There was considerable variation in the type of errors that the various systems detected; the failure rate for the plans with errors varied between 0% and 72%. When using 2% and 2 mm criteria and 95% as a pass rate the Delta(4®) detected 15 of 20 errors, OCTAVIUS(®) detected 8 of 20 errors, COMPASS detected 8 of 20 errors, and Epiqa™ detected 20 of 20 errors. It was also found that the calibration and measuring procedure could benefit from improvements for some of the patient QA systems. CONCLUSIONS The various systems can detect various errors and the sensitivity to the introduced errors depends on the plan. There was poor correlation between the gamma evaluation pass rates of the QA procedures and the deviations observed in the dose volume histograms.
منابع مشابه
A Specific Patient Quality Assurance (PSQA) procedure for a Co-60 source based High Dose Rate Brachytherapy
Introduction: In radiation therapy, accurate dose determination and precise dose delivery to the tumor are directly associated with better treatment outcomes in terms of higher tumor control and lower post radiation therapy complications. The current study aims the development and clinical application of the Patient Specific Quality Assurance (PSQA) procedures for nasopharyngea...
متن کاملEvaluation of the sensitivity of two 3D diode array dosimetry systems to setup error for quality assurance (QA) of volumetric‐modulated arc therapy (VMAT)
The purpose of this study is to evaluate the sensitivities of 3D diode arrays to setup error for patient-specific quality assurance (QA) of volumetric-modulated arc therapy (VMAT). Translational setup errors of ± 1, ± 2, and ± 3 mm in the RL, SI, and AP directions and rotational setup errors of ± 1° and ± 2° in the pitch, roll, and yaw directions were set up in two phantom systems, ArcCHECK and...
متن کاملA comprehensive quality assurance program for personnel and procedures in radiation oncology: value of voluntary error reporting and checklists.
PURPOSE This report describes the value of a voluntary error reporting system and the impact of a series of quality assurance (QA) measures including checklists and timeouts on reported error rates in patients receiving radiation therapy. METHODS AND MATERIALS A voluntary error reporting system was instituted with the goal of recording errors, analyzing their clinical impact, and guiding the ...
متن کاملThe sensitivity of gamma-index method to the positioning errors of high-definition MLC in patient-specific VMAT QA for SBRT
BACKGROUND To investigate the sensitivity of various gamma criteria used in the gamma-index method for patient-specific volumetric modulated arc therapy (VMAT) quality assurance (QA) for stereotactic body radiation therapy (SBRT) using a flattening filter free (FFF) photon beam. METHODS Three types of intentional misalignments were introduced to original high-definition multi-leaf collimator ...
متن کاملUsing a photographic image processing method as a part of the QA program in some radiotherapy departments where CTs are involved with COVID-19 patients
Background & Aims: Owing to the significant role of CT images in diagnosis and follow up of patients affected with coronavirus, the imaging section of cancer centers in some countries engaged in providing services to COVID-19 patients. The aim of this study was to introduce a CT-independent photographic-based QA method in some radiotherapy departments where CTs are involved with COVID-19 patien...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Medical physics
دوره 40 3 شماره
صفحات -
تاریخ انتشار 2013