چکیده

BACKGROUND Decreasing the stiffness of locked plating constructs can promote natural fracture healing by controlled dynamization of the fracture. This biomechanical study compared the effect of 4 different stiffness reduction methods on interfragmentary motion by measuring axial motion and shear motion at the fracture site. METHODS Distal femur locking plates were applied to bridge a metadiaphyseal fracture in femur surrogates. A locked construct with a short-bridge span served as the nondynamized control group (LOCKED). Four different methods for stiffness reduction were evaluated: replacing diaphyseal locking screws with nonlocked screws (NONLOCKED); bridge dynamization (BRIDGE) with 2 empty screw holes proximal to the fracture; screw dynamization with far cortical locking (FCL) screws; and plate dynamization with active locking plates (ACTIVE). Construct stiffness, axial motion, and shear motion at the fracture site were measured to characterize each dynamization methods. RESULTS Compared with LOCKED control constructs, NONLOCKED constructs had a similar stiffness (P = 0.08), axial motion (P = 0.07), and shear motion (P = 0.97). BRIDGE constructs reduced stiffness by 45% compared with LOCKED constructs (P < 0.001), but interfragmentary motion was dominated by shear. Compared with LOCKED constructs, FCL and ACTIVE constructs reduced stiffness by 62% (P < 0.001) and 75% (P < 0.001), respectively, and significantly increased axial motion, but not shear motion. CONCLUSIONS In a surrogate model of a distal femur fracture, replacing locked with nonlocked diaphyseal screws does not significantly decrease construct stiffness and does not enhance interfragmentary motion. A longer bridge span primarily increases shear motion, not axial motion. The use of FCL screws or active plating delivers axial dynamization without introducing shear motion.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Dynamic Stabilization with Active Locking Plates Delivers Faster, Stronger, and More Symmetric Fracture-Healing.

BACKGROUND Axial dynamization of fractures can promote healing, and overly stiff fixation can suppress healing. A novel technology, termed active plating, provides controlled axial dynamization by the elastic suspension of locking holes within the plate. This prospective, controlled animal study evaluated the effect of active plates on fracture-healing in an established ovine osteotomy model. W...

متن کامل

Functional Results of Unstable (Type 2) Distal Clavicle Fractures Treated with Superior Anterior Locking Plate

Background: The treatment of distal clavicle fracture is always a challenge, as it is mostly unstable and has higherrate of delayed union, malunion, non-union and associated acromioclavicular arthritis. So the management of thesefractures remains controversial. The purpose of this study is to evaluate the functional results of Type 2 distal endclavicle fractures treated with superior anterior l...

متن کامل

Supracondylar Osteotomy in Valgus Knee: Angle Blade Plate Versus Locking Compression Plate

  Background: There are few studies comparing the biomechanical properties of angled blade plate and locking compression plates in supracondylar osteotomy. In the current randomized study, we prospectively compared the clinical and radiological outcomes of supracondylar osteotomy using these two plates. Methods: Forty patients with valgus knee malalignment were randomly assigned to two equal nu...

متن کامل

The Outcome of Proximal Humeral Locking Plates in the Management of Three and Four Part Proximal Humeral Fractures in Special Cohort of Young Patients in High Velocity Trauma

Background: Locking plate fixation provides satisfactory outcome following proximal humerus fractures. None of theprevious studies selectively evaluate the outcome in young patients. This study evaluates outcome of locking platesystem in the treatment of acute three- and four-part fractures in young patients.Methods: In this prospective study we included all patients who were less than 60 years...

متن کامل

Internal fixation of dorsally displaced fractures of the distal part of the radius. A biomechanical analysis of volar plate fracture stability.

BACKGROUND Volar plate fixation with use of either a locking plate or a neutralization plate has become increasingly popular among surgeons for the treatment of dorsally comminuted extra-articular distal radial fractures. The purpose of the present study was to compare the relative stability of five distal radial plates (four volar and one dorsal), all of which are commonly used for the treatme...