Programmable CSF shunt valve: in vitro assessment of MR imaging safety at 3T.

نویسندگان

  • F G Shellock
  • R Habibi
  • J Knebel
چکیده

A programmable CSF shunt valve was assessed for magnetic field interactions, heating (transmit-receive body radio-frequency coil; whole-body averaged specific absorption rate, 2.1 W/kg), functional alterations, and artifacts at 3T. The programmable valve showed minor magnetic field interactions and heating was not excessive (+0.8 degrees C). The function of the programmable valve was not altered by multiple exposures to the 3T scanner or from exposure to various MR imaging conditions. Therefore, this implant is safe for a patient undergoing MR imaging at 3T or less when the radiologist follows specific safety guidelines. Artifacts for the programmable valve were relatively large in relation to the size and shape of the valve; this finding may impact the diagnostic use of MR imaging if the area of interest is in proximity to this implant.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Programmable CSF shunt valves: radiographic identification and interpretation.

The programmable CSF shunt valve has become an important tool in hydrocephalus treatment, particularly in the NPH population and in pediatric patients with complex hydrocephalus. The purpose of this study is to provide a single reference for the identification of programmable shunt valves and the interpretation of programmable shunt valve settings. Four major manufacturers of programmable shunt...

متن کامل

Magnetically programmable shunt valve: MRI at 3-Tesla.

A magnetically programmable cerebrospinal fluid (CSF) shunt valve (Codman Hakim Programmable Valve, Codman, a Johnson & Johnson Company, Raynham, MA) was assessed for magnetic field interactions, heating, artifacts and functional changes at 3-Tesla. The programmable valve showed minor magnetic field interactions and heating (+0.4 degrees C). Artifacts were relatively large in relation to the si...

متن کامل

Flow-sensitive MR imaging of ventriculoperitoneal shunts: in vitro findings, clinical applications, and pitfalls.

To determine ventriculoperitoneal shunt patency during routine MR imaging of the head, 23 patients were studied with T1-weighted fast-field-echo scans. Without knowledge of the results of previous MR/CT studies or of the patients' clinical history, we reviewed the fast-field-echo studies and divided them according to those judged to have shunt flow (18) and those judged not to have flow (five)....

متن کامل

Maladjustment of programmable ventricular shunt valves by inadvertent exposure to a common hospital device

Background Programmable ventricular shunt valves are commonly used to treat hydrocephalus. They can be adjusted to allow for varying amounts of cerebrospinal fluid (CSF) flow using an external magnetic programming device, and are susceptible to maladjustment from inadvertent exposure to magnetic fields. Case Description We describe the case of a 3-month-old girl treated for hydrocephalus with...

متن کامل

Normal findings on brain fluid-attenuated inversion recovery MR images at 3T.

BACKGROUND AND PURPOSE Fluid attenuated inversion recovery (FLAIR) MR imaging of the brain has become a routine tool for assessing lesions in patients with suspected neurologic disorders. There is growing interest in 3T brain FLAIR MR imaging but little normative data are available. The purpose of this study was to evaluate the frequency and topography of cerebral hyperintensities seen with FLA...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • AJNR. American journal of neuroradiology

دوره 27 3  شماره 

صفحات  -

تاریخ انتشار 2006