Reliability and limits of agreement of circumferential, water displacement, and optoelectronic volumetry in the measurement of upper limb lymphedema.
نویسندگان
چکیده
We conducted a reliability comparison study to determine the intrarater and inter-rater reliability and the limits of agreement of the volume estimated by circumferential measurements using the frustum sign method and the disk model method, by water displacement volumetry, and by infrared optoelectronic volumetry in the assessment of upper limb lymphedema. Thirty women with lymphedema following axillary lymph node dissection surgery for breast cancer surgery were enrolled. In each patient, the volumes of the upper limbs were estimated by three physical therapists using circumference measurements, water displacement and optoelectronic volumetry. One of the physical therapists performed each measure twice. Intraclass correlation coefficients (ICCs), relative differences, and limits of agreement were determined. Intrarater and interrater reliability ICCs ranged from 0.94 to 1. Intrarater relative differences were 1.9% for the disk model method, 3.2% for the frustum sign model method, 2.9% for water displacement volumetry, and 1.5% for optoelectronic volumetry. Intrarater reliability was always better than interrater, except for the optoelectronic method. Intrarater and interrater limits of agreement were calculated for each technique. The disk model method and optoelectronic volumetry had better reliability than the frustum sign method and water displacement volumetry, which is usually considered to be the gold standard. In terms of low-cost, simplicity, and reliability, we recommend the disk model method as the method of choice in clinical practice. Since intrarater reliability was always better than interrater reliability (except for optoelectronic volumetry), patients should therefore, ideally, always be evaluated by the same therapist. Additionally, the limits of agreement must be taken into account when determining the response of a patient to treatment.
منابع مشابه
Reliability of the Inverse Water Volumetry Method to Measure the Volume of the Upper Limb.
BACKGROUND Lymphedema of the upper extremity is a common side effect of lymph node dissection or irradiation of the axilla. Several techniques are being applied in order to examine the presence and severity of lymphedema. Measurement of circumference of the upper extremity is most frequently performed. An alternative is the water-displacement method. The aim of this study was to determine the r...
متن کاملComparative study between mobile infrared optoelectronic volumetry with a Perometer and two commonly used methods for the evaluation of arm volume in patients with breast cancer related lymphedema of the arm.
There is no consensus on the definition of Breast Cancer Related Lymphedema of the arm (BCRL) because there are no agreed standards in measurement methods and diagnostic criteria. The main objective of this study is to compare mobile infrared optoelectronic volumetry with a Perometer with two commonly used methods for the evaluation of arm volume in patients with different degrees of BCRL. Bila...
متن کاملCorrelation between bioelectrical spectroscopy and perometry in assessment of upper extremity swelling.
Lymphedema is a common side effect of breast cancer treatment and is associated with increased upper extremity volume, functional impairment, and pain. While there is no cure for lymphedema, physical therapy treatment can often alleviate symptoms. To measure the efficacy of treatment, accurate assessment of the limbs is important. Current methods of assessment are complex (water displacement), ...
متن کاملValidity and intra- and interobserver reliability of an indirect volume measurements in patients with upper extremity lymphedema.
We investigated a method of indirect volume measurement that utilized surface measurements and a simplified formula derived from the formula for a frustum (Sitzia's method) to determine limb volumes in patients with breast cancer-related lymphedema of the upper extremity. Repeated measurements of upper-extremity limb volume were obtained by two observers on both upper extremities of 30 women wi...
متن کاملHand volume estimates based on a geometric algorithm in comparison to water displacement.
Assessing changes in upper extremity limb volume during lymphedema therapy is important for determining treatment efficacy and documenting outcomes. Although arm volumes may be determined by tape measure, the suitability of circumference measurements to estimate hand volumes is questionable because of the deviation in circularity of hand shape. Our aim was to develop an alternative measurement ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Lymphology
دوره 40 1 شماره
صفحات -
تاریخ انتشار 2007