The Sysmex CS-5100 coagulation analyzer offers comparable analytical performance and excellent throughput capabilities
نویسندگان
چکیده
OBJECTIVES This study compared the new high-volume blood coagulation analyzer Sysmex CS-5100 System™ (Siemens Healthcare Diagnostics, Erlangen, Germany) to the mid-volume blood coagulation analyzer Sysmex CS-2000i System™ (Siemens) for analytical performance. Additionally, the operational performance of the Sysmex CS-5100 System was compared with the blood coagulation analyzer ACL TOP 700 (Instrumentation Laboratory, Werfen Group, Kirchheim bei Munchen, Germany). MATERIALS AND METHODS We compared the Sysmex CS-5100 to the Sysmex CS-2000i and the ACL TOP analyzer for routine coagulation, chromogenic and immunological assays. Imprecision studies were performed for the Sysmex CS-5100 and Sysmex CS-2000i systems. A throughput and STAT analysis comparison of the CS-5100 and the ACL TOP was performed. A stress test was performed to characterize the robustness and the error rate of the CS-5100. We also performed correlation analysis between the CS-5100 and the CS-2000i or the ACL TOP in the measurement of patients' samples. RESULTS The inter-assay precision using the CS systems was impressive (inter-assay CV generally <3.5%) and the correlation between the two Sysmex analyzers was excellent. In the throughput study, the CS-5100 completed the measurement of 100 samples (210 results) in less than 49 min. CONCLUSIONS Our results demonstrated that the CS-5100 is a robust high-throughput analyzer, well-suited for coagulation laboratories.
منابع مشابه
Evaluation of the Diagnostic Performance of theSysmex XT-2000i Automated Hematology Analyzerin the Detection of Immature Granulocyte
Immature Granulocytes (IG) are normally not present in peripheral blood. Their presence is an important indicator of enhanced bone marrow activation. The present study evaluated the diagnostic performance of the Sysmex XT-2000i automated hematology analyzer in the detection of myelocytes and/or metamyelocytes in patients affected by non-haematological disease, in comparison with manual microsco...
متن کاملPerformance evaluation and multicentre study of a von Willebrand factor activity assay based on GPIb binding in the absence of ristocetin.
The functional activity of von Willebrand factor (VWF) is most frequently measured by using the ristocetin cofactor assay (VWF:RCo). However, the method's drawbacks include unsatisfactory precision, sensitivity and availability of automated system applications. We have developed an alternative assay (INNOVANCE VWF Ac) that is based on the binding of VWF to recombinant glycoprotein Ib (GPIb). Tw...
متن کاملImproving the efficiency and efficacy of pre-analytical and analytical work-flow of urine cultures with urinary flow cytometry.
Quantitative urine culture is one of the most frequently requested tests in microbiology laboratories. An automated system for screening purposes is strongly needed to save technical staff time and obtain rapid results. Our study investigated 1.047 urine samples collected from inpatients and outpatients with a commercial vacutainer system (Becton Dickinson, Milan, Italy) and compared a second-g...
متن کاملFully automated immunoassay for quantitative determination of FXIII.
UNLABELLED Coagulation factor XIII (FXIII) is essential for clot stabilization. Deficiency of FXIII is associated with a risk of bleeding and impaired wound healing. Substitution therapy with FXIII remedies for patients with low plasma levels of FXIII requires diagnostic quantification of the factor before and during therapy. Here, we describe a prototype of a preliminary research immunoassay f...
متن کاملNew fluorescent method (PLT-F) on Sysmex XN2000 hematology analyzer achieved higher accuracy in low platelet counting.
OBJECTIVES In thrombocytopenia, high accuracy and precision of low platelet count is essential for appropriate decisions. The recently introduced Sysmex XN2000 analyzer (Sysmex, Kobe, Japan) offers 3 methods for platelet counting: impedance (PLT-I), optical (PLT-O), and a new fluorescence method (PLT-F). The precision of the PLT-F method in blood samples with platelet counts less than 50 × 10(3...