High-energy-resolution X-ray monochromator calibration using the detailed-balance principle
نویسندگان
چکیده
A new method is presented to calibrate an X-ray energy scale with sub-meV relative accuracy by using the detailed-balance principle of the phonon creation and annihilation. This method is conveniently used to define or verify the energy scale of high-energy-resolution monochromators that are used in inelastic X-ray scattering and nuclear resonant inelastic X-ray scattering instruments at synchrotron radiation facilities. This method does not rely on sample properties and its precision only depends on the statistical data quality. Well calibrated instruments are essential for reliable comparison of data sets obtained at different synchrotron radiation beamlines, of data with theoretical predictions, and of data from other techniques such as neutron or light scattering. The principle of the detailed-balance method is described in this paper and demonstrated experimentally.
منابع مشابه
Ultra-stable sub-meV monochromator for hard X-rays
A high-resolution silicon monochromator suitable for 21.541 keV synchrotron radiation is presented that produces a bandwidth of 0.27 meV. The operating energy corresponds to a nuclear transition in (151)Eu. The first-of-its-kind, fully cryogenic design achieves an energy-alignment stability of 0.017 meV r.m.s. per day, or a 100-fold improvement over other meV-monochromators, and can tolerate hi...
متن کاملHigh Resolution Krypton M4.5 X-ray Emission Spectra
High resolution M,,, (3d -+ 4p) x-ray emission spectra from a kwton plasma were measured using a recently developed grazing-incidence reflectiongrating monochromator/spectrometer with very high flux rates at extreme ultraviolet and soft x-ray wavelengths. The nominal resolving power of the instrument, E/AE, is about 300 in this energy range (-80 eV). Three dipole-allowed 3d -+ 4p emission lines...
متن کاملProfiled bar transmission gratings: soft-x-ray calibration of new Kirchoff solutions.
A new analytical model, derived rigorously from scalar diffraction theory, accurately fits soft-x-ray measurements of symmetrical profile gold transmission gratings in all diffracted orders. The calibration system selects numerous photon energies by use of a high-resolution grazing-incidence monochromator and a dc e-beam source. Fine-period free-standing gratings exhibit limited performance and...
متن کاملA diffracted-beam monochromator for long linear detectors in X-ray diffractometers with Bragg-Brentano parafocusing geometry.
A new diffracted-beam monochromator has been developed for Bragg-Brentano X-ray diffractometers equipped with a linear detector. The monochromator consists of a cone-shaped graphite highly oriented pyrolytic graphite crystal oriented out of the equatorial plane such that the parafocusing geometry is preserved over the whole opening angle of the linear detector. In our standard setup a maximum w...
متن کاملA laboratory-based hard x-ray monochromator for high-resolution x-ray emission spectroscopy and x-ray absorption near edge structure measurements.
We report the development of a laboratory-based Rowland-circle monochromator that incorporates a low power x-ray (bremsstrahlung) tube source, a spherically bent crystal analyzer, and an energy-resolving solid-state detector. This relatively inexpensive, introductory level instrument achieves 1-eV energy resolution for photon energies of ∼5 keV to ∼10 keV while also demonstrating a net efficien...
متن کامل