Hemodynamic Ramp Tests in Patients With Left Ventricular Assist Devices.

نویسندگان

  • Nir Uriel
  • Gabriel Sayer
  • Karima Addetia
  • Savitri Fedson
  • Gene H Kim
  • Daniel Rodgers
  • Eric Kruse
  • Keith Collins
  • Sirtaz Adatya
  • Nitasha Sarswat
  • Ulrich P Jorde
  • Colleen Juricek
  • Takeyoshi Ota
  • Valluvan Jeevanandam
  • Daniel Burkhoff
  • Roberto M Lang
چکیده

OBJECTIVES This study tested whether combined invasive hemodynamic and echocardiographic ramp tests can help optimize patient management. BACKGROUND Guidelines for optimizing speed and medications in continuous flow ventricular assist device (cfLVAD) patients are mainly based on expert opinion. METHODS Thirty-five cfLVAD patients (21 HeartMate II [Thoratec, Pleasanton, California] and 14 HVAD [HeartWare International, Framingham, Massachusetts]) underwent ramp tests with right heart catheterization (including central venous pressure [CVP], pulmonary artery pressure, pulmonary capillary wedge pressure [PCWP], and blood pressure) and echocardiography. Data were recorded at up to 9 speed settings. Speed changes were in steps of 400 revolutions per minute (RPM) for HeartMate II (8,000 to 12,000 RPM) and 100 RPM for HVAD (2,300 to 3,200 RPM) patients. RESULTS Only 42.9% of patients had normal CVPs and PCWPs at their original RPM settings. Going from lowest to highest speeds, cardiac output improved by 0.16 ± 0.19 l/min/step (total change 1.28 ± 1.41 l/min) and PCWP decreased by 1.23 ± 0.85 mm Hg/step (total change 9.9 ± 6.5 mm Hg). CVP and systolic blood pressure did not change significantly with RPM. RPM were adjusted based on test results to achieve CVPs and PCWPs as close to normal limits as possible, which was feasible in 56% of patients. For the remainder, results indicated which type of medical management should be pursued. CONCLUSIONS Use of combined hemodynamic and echocardiographic ramp tests in patients provides objective means of optimizing RPM, and has the potential to guide medical management. It remains to be tested whether this strategy has a beneficial impact on quality of life or clinical outcomes.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Left Ventricular Decompression During Speed Optimization Ramps in Patients Supported by Continuous-Flow Left Ventricular Assist Devices: Device-Specific Performance Characteristics and Impact on Diagnostic Algorithms.

BACKGROUND Echocardiographic ramp tests have been widely used to help guide speed adjustments and for identification of potential device malfunctions in patients with axial continuous-flow left ventricular assist devices (LVADs) (Heartmate II LVAD [HMII]). Recently, the use of centrifugal-flow LVADs (Heartware LVAD [HVAD]) has been on the rise. The purpose of this study was to evaluate the util...

متن کامل

Design an Equivalent Left Ventricular Assist Device for Medical Equipment Labs

LVAD is a mechanical pump supporting a weak heart function and blood flow. Sometimes, the heart may not recover fast enough to take over the pumping action immediately after surgery, in such patients a temporary support device has been employed to maintain the pumping action until the patient’s own heart recovers. This device can be considered as a temporary alternative before the process of ar...

متن کامل

Atypical tamponade hemodynamic in a patient with temporary left ventricular assist device.

Postoperative bleeding and tamponade are considered major complications after implantation of left ventricular assist devices. Firstly, 40% of patients supported by ventricular assist devices experience early postoperative bleeding, and 20% developed tamponade requiring re-exploration. Secondly, we present a case of a patient with tamponade, on temporary left ventricular assist device (LVAD) su...

متن کامل

Echocardiography and Continuous-Flow Left Ventricular Assist Devices: Evidence and Limitations.

Echocardiography is the most used imaging modality in the growing population of patients with advanced heart failure undergoing continuous-flow, durable mechanical circulatory support. However, no guidelines for the use of echocardiography in this population exist, evidence for core applications is limited and conflicting, and newer centrifugal-flow devices have been subject to minimal study. A...

متن کامل

Simultaneous ramp right heart catheterization and echocardiography in a ReliantHeart left ventricular assist device

Many clinicians caring for patients with continuous flow left ventricular assist devices (CF-LVAD) use ramp right heart catheterization (RHC) studies to optimize pump speed and also to troubleshoot CF-LVAD malfunction. An investigational device, the ReliantHeart Heart Assist 5 (Houston, TX), provides the added benefit of an ultrasonic flow probe on the outflow graft that directly measures flow ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • JACC. Heart failure

دوره 4 3  شماره 

صفحات  -