Medication nebulizer performance. Effects of diluent volume, nebulizer flow, and nebulizer brand.

نویسندگان

  • D Hess
  • D Fisher
  • P Williams
  • S Pooler
  • R M Kacmarek
چکیده

BACKGROUND Medication nebulizers are commonly used to delivery aerosolized medications to patients with respiratory disease. We evaluated output and respirable aerosol available to the patient (inhaled mass) for 17 medication nebulizers using a spontaneous breathing lung model. METHODS Three nebulizer fill volumes (3, 4, and 5 mL containing 2.5 mg of albuterol) and 3 oxygen flows (6, 8, and 10 L/min) were evaluated using the 17 nebulizers. A cotton plug at the nebulizer mouthpiece was used to trap aerosol during simulated spontaneous breathing. Following each trial, the amount of albuterol remaining in the nebulizer and the amount deposited in the cotton plug were determined spectrophotometrically. Aerosol particle size was determined using an 11-stage cascade impactor. RESULTS Increasing fill volume decreased the amount of albuterol trapped in the dead volume (p < 0.001) and increased the amount delivered to the patient (p < 0.001). Increasing flow increased the mass output of particles in the respirable range of 1 to 5 microns (p = 0.004), but the respirable mass delivered to the patient was affected to a greater extent by nebulizer brand (p < 0.001) than flow. Although 2.5 mg of albuterol was placed into the nebulizers, less than 0.5 mg in the respirable range of 1 to 5 microns was delivered to the mouthpiece. CONCLUSIONS The performance of medication nebulizers is affected by fill volume, flow, and nebulizer brand. When they are used for research applications, the nebulizer characteristics must be evaluated and reported for the conditions used in the investigation.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

In-Vitro Comparison of Six Continuous Nebulizers*

BACKGROUND: Continuous nebulization has become popular for treating patients with severe reversible airway disease. Continuous nebulization therapy has been shown to be safe, effective, and labor efficient as compared to other traditional medication aerosol treatments 1-11 . Although numerous studies have been done evaluating the clinically efficacy of continuous nebulization, little work has b...

متن کامل

Comparison of twenty three nebulizer/compressor combinations for domiciliary use.

We have assessed the physical and dynamic characteristics of 23 home jet nebulizer/compressor combinations currently available in the UK and Europe. The combinations were evaluated in terms of pressure-flow characteristics, aerosol mass distribution, volume output, electrical costs, and sound level. In addition, we determined the effect of nebulizer fill volume on aerosol mass distribution and ...

متن کامل

A comparison of responses to albuterol delivered by two aerosol devices.

Nineteen outpatients with stable obstructive pulmonary disease (mean forced expiratory volume in one second [FEV1], 1.00 + 0.10 L) were evaluated for airway response to albuterol (salbutamol) administered by metered-dose inhaler and Bosch ultrasonic nebulizer (BUSN). Albuterol administered by metered-dose inhaler but not by nebulizer caused a significant increase in FEV1 and the mean forced exp...

متن کامل

Comparison of a Nebulizer Using a Novel Aerosol Generator with a Standard Ultrasonic Nebulizer Designed for Use during Mechanical Ventilation

Figure 1: Diagram of SUN 145 (from product instruction manual) Ultrasonic nebulizers (USNs) generate aerosols without changing ventilator volumes or alarms. However, USN output varies with gas flow, is position dependent and generates heat. We wanted to determine whether the AeronebTM Professional Nebulizer System (Aeroneb Pro; Aerogen, Inc., Sunnyvale, CA)* has similar limitations. METHODS: We...

متن کامل

Medical nebulizer performance: effects of cascade impactor temperature.

BACKGROUND During operation of a jet nebulizer, the temperature of the nebulizer outlet could decrease by more than 10 degrees C, depending on the nebulizer type and operating conditions, such as driving flow rate and fill volume. The droplet size distribution generated from the nebulizer can be measured by a cascade impactor. However, when the cascade impactor is operated at ambient room tempe...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Chest

دوره 110 2  شماره 

صفحات  -

تاریخ انتشار 1996