Monte Carlo calculations of beam quality correction factors kQ for electron dosimetry with a parallel-plate Roos chamber.
نویسندگان
چکیده
Current dosimetry protocols (AAPM, IAEA, DIN) recommend the use of parallel-plate ionization chambers for the measurement of absorbed dose-to-water in clinical electron beams. For well-guarded plane-parallel chambers, it is assumed that the perturbation correction pQ is unity for all electron energies. In this study, we present detailed Monte Carlo simulations with the EGSnrc code for the widely used Roos parallel-plate chamber which is, besides other plane-parallel chamber types, recommended in all protocols. We have calculated the perturbation corrections pcav and pwall for a wide range of electron energies and for 60Co. While our results confirm the recommended value of unity for the cavity perturbation pcav, the wall-correction factor pwall depends on electron energy and decreases with increasing electron energy. For the lowest electron energies in this study (R50 approximately 2 cm), pwall deviates from unity by up to 1.5%. Using the perturbation factors for the different electron energies and those for the reference beam quality, 60Co, we have calculated the beam quality correction factor kQ. For electron energies E0>9 MeV (R50>4 cm), the calculated values are in good agreement with the data published in the IAEA protocol. Deviations in the range of 0.5-0.8% are found for R50<3 cm.
منابع مشابه
Monte Carlo calculations of beam quality correction factors
Current dosimetry protocols (AAPM, IAEA, DIN) recommend the use of parallel-plate ionization chambers for the measurement of absorbed dose-towater in clinical electron beams. For well-guarded plane-parallel chambers, it is assumed that the perturbation correction pQ is unity for all electron energies. In this study, we present detailed Monte Carlo simulations with the EGSnrc code for the widely...
متن کاملA Comparison of dosimetric parameters between IAEA TRS-398, AAPM TG-51 protocols and Monte-Carlo simulation
Background: Two protocols of AAPM TG-51 and IAEA TRS-398 were compared followed by a measurement and Monte Carlo simulation of beam quality correction factor, KQ, AAPM TG-51 and IAEA TRS-398 protocols were compared for the absorbed dose to water for DW, and KQ parameters. Materials and Methods: Dose measurements by either protocols were performed with cylindrical and plane parallel ch...
متن کاملMonte Carlo calculations of electron beam quality conversion factors for several ion chamber types.
PURPOSE To provide a comprehensive investigation of electron beam reference dosimetry using Monte Carlo simulations of the response of 10 plane-parallel and 18 cylindrical ion chamber types. Specific emphasis is placed on the determination of the optimal shift of the chambers' effective point of measurement (EPOM) and beam quality conversion factors. METHODS The EGSnrc system is used for calc...
متن کاملMonte Carlo calculations of correction factors for plane-parallel ionization chambers in clinical electron dosimetry.
Recent standard dosimetry protocols recommend that plane-parallel ionization chambers be used in the measurements of depth-dose distributions or the calibration of low-energy electron beams with beam quality R50 <4 g/cm2. In electron dosimetry protocols with the plane-parallel chambers, the wall correction factor, Pwall, in water is assumed to be unity and the replacement correction factor, Pre...
متن کاملSmall photon field dosimetry using EBT2 Gafchromic film and Monte Carlo simulation
Background: Small photon fields are increasingly used in modern radiotherapy especially in intensity modulated radiation therapy (IMRT) and stereotactic radiosurgery (SRS) treatments. Accurate beam profile and central axis depth doses measurements of such beams are complicated due to the electron disequilibrium. Hence the EBT2 (external beam therapy) Gafchromic film was used for dosimetry of sm...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Physics in medicine and biology
دوره 53 6 شماره
صفحات -
تاریخ انتشار 2008