Quantitative imaging of bremsstrahlung from pure beta emitters is proposed as a means for in vivo management of antibody therapy. The method involves the use of high-energy collimation, an empirically selected broad photon energy window to enhance detector sensitivity, and a Wiener restoration filter to compensate for system blur. The measured and filtered data were obtained for an idealized scattering medium and isolated spherical sources. An effective linear attenuation coefficient of about 0.13 cm-1 was determined from the raw image data of 32P. A coefficient of 0.14 cm-1 was determined after the images were restored using the Wiener filter. The measured attenuation was not significantly dependent on the size of the region of interest or the size of the source. Its variation was within the experimental error of measurement (+/- 5%). The measured sensitivity (6 x 10(-6) cps/Bq) was sufficient for imaging therapy doses of 32P or 90Y.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Bremsstrahlung imaging from the liver using the Monte Carlo simulation

Introduction: Most beta and gamma radiation radioisotopes used for treatment are not suitable for imaging. The bremsstrahlung images on a conventional gamma camera helped to localize the radionuclide within and outside of the lesion. Secondary scattering of gamma rays of higher energy and bremsstrahlung causes contamination in the energy window and reducing the contrast and resolution of the im...

متن کامل

Quantitative SPECT and planar 32P bremsstrahlung imaging for dosimetry purpose –An experimental phantom study

Background: In this study, Quantitative 32P bremsstrahlung planar and SPECT imaging and consequent dose assessment were carried out as a comprehensive phantom study to define an appropriate method for accurate Dosimetry in clinical practice. Materials and Methods: CT, planar and SPECT bremsstrahlung images of Jaszczak phantom containing a known activity of 32P were acquired. In addition, Phanto...

متن کامل

ارزیابی کولیماتورهای لایه‌ای در تصویربرداری تابش ترمزی حین رادیو دارو درمانی با فسفر-32

Background and purpose: Imaging which is used in nuclear medicine is done using emit gamma rays whereas therapy in nuclear medicine is performed by applying radionuclide’s which emit beta radiation. Using Bremsstrahlung is a new field which is recently introduced in imaging. In the Bremsstrahlung imaging, collimator design should not be based on energy peak and whole energy spectrum should be c...

متن کامل

A camera based calculation of 99mTc-MAG3 clearance using conjugate views method [Persian]

Background: Measurement of absolute or differential renal function using radiotracers plays an important role in the clinical management of various renal diseases. Gamma camera quantitative methods in approximation of renal clearance may potentially be as accurate as plasma clearance methods. However some critical factors such as kidney depth and background counts are still troublesome in...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of nuclear medicine : official publication, Society of Nuclear Medicine

دوره 33 1  شماره 

صفحات  -

تاریخ انتشار 1992