Effective dose from direct and indirect digital panoramic units
نویسندگان
چکیده
PURPOSE This study aimed to provide comparative measurements of the effective dose from direct and indirect digital panoramic units according to phantoms and exposure parameters. MATERIALS AND METHODS Dose measurements were carried out using a head phantom representing an average man (175 cm tall, 73.5 kg male) and a limbless whole body phantom representing an average woman (155 cm tall, 50 kg female). Lithium fluoride thermoluminescent dosimeter (TLD) chips were used for the dosimeter. Two direct and 2 indirect digital panoramic units were evaluated in this study. Effective doses were derived using 2007 International Commission on Radiological Protection (ICRP) recommendations. RESULTS The effective doses of the 4 digital panoramic units ranged between 8.9 µSv and 37.8 µSv. By using the head phantom, the effective doses from the direct digital panoramic units (37.8 µSv, 27.6 µSv) were higher than those from the indirect units (8.9 µSv, 15.9 µSv). The same panoramic unit showed the difference in effective doses according to the gender of the phantom, numbers and locations of TLDs, and kVp. CONCLUSION To reasonably assess the radiation risk from various dental radiographic units, the effective doses should be obtained with the same numbers and locations of TLDs, and with standard hospital exposure. After that, it is necessary to survey the effective doses from various dental radiographic units according to the gender with the corresponding phantom.
منابع مشابه
Using GafChromic film to estimate the effective dose from dental cone beam CT and panoramic radiography.
OBJECTIVES To demonstrate the feasibility of GafChromic(®) XR-QA2 (ISP Corp., Wayne, NJ) as a dosemeter when performing measurements of the effective dose from three cone beam CT (CBCT) units and to compare the doses from examinations of three common dental clinical situations. A second aim was to compare the radiation doses for three digital panoramic units with the doses for the CBCT units. ...
متن کاملQuality comparison of direct digital panoramic radiography and computed radiography panoramic
BACKGROUND AND AIM: Digital panoramic X-ray images can be captured using photostimulable phosphors or solid-state detectors (i.e. charge-coupled devices and Flat-Panels). The first category is defined as computed radiography (CR) or semi-direct radiography. The second technology that uses solid-state detectors is known as direct digital radiography (DDR). Both of these technologies have their o...
متن کاملDiagnostic Reference Level Arising from Dental Panoramic Radiography
Introduction The present work describes a study in which, based on patient dose measurements, thermoluminescent dosimeters were used to obtain the diagnostic reference level arising from panoramic radiography. Materials and Methods Ten panoramic units and a sample of 15 patients per X-ray unit were studied. Two thermoluminescent dosimeter chips were placed on the skin surface of selected organs...
متن کاملDose-area product measurements in panoramic dental radiology.
In this study, dose-area product (DAP) measurements in panoramic dental radiology have been performed in Germany. The results obtained in this study were proposed as diagnostic reference levels (DRLs). A representative number of dental panoramic units, both with digital and conventional image receptors, have been chosen. Common statistical parameters such as mean, standard deviation and 3rd qua...
متن کاملMethods for monitoring patient dose in dental radiology.
Different types of X-ray equipment are used in dental radiology, such as intra-oral, panoramic, cephalometric, cone-beam computed tomography (CBCT) and multi-slice computed tomography (MSCT) units. Digital receptors have replaced film and screen-film systems and other technical developments have been made. The radiation doses arising from different types of examination are sparsely documented a...
متن کامل