Experimental evaluation of the Spiegelberg intracranial pressure and intracranial compliance monitor. Technical note.

نویسندگان

  • Y H Yau
  • I R Piper
  • R E Clutton
  • I R Whittle
چکیده

The goal of this study was to compare the Spiegelberg intraventricular intracranial pressure (ICP)/intracranial compliance monitoring device, which features an air-pouch balloon catheter, with existing gold-standard methods of measuring ICP and intracranial compliance. A Spiegelberg intraventricular catheter, a standard intraventricular catheter, and a Codman intraparenchymal ICP microsensor were placed in five sheep, which previously had been given anesthetic and paralytic agents, to allow comparative measurement of ICP at incremental levels (range 5-50 mm Hg). Intracranial pressure measured using the Spiegelberg intraventricular air-pouch balloon catheter displayed a linear correlation with ICP measured using the standard intraventricular fluid-filled catheter (r2 = 0.9846, p < 0.001; average bias -0.74 mm Hg), as well as with ICP measured using the Codman intraparenchymal strain-gauge sensor (r2 = 0.9778, p < 0.001; average bias 0.01 mm Hg). Automated measurements of intraventricular compliance obtained using the Spiegelberg compliance device were compared with compliance measurements that were made using the gold-standard manual cerebrospinal fluid bolus injection technique at ICPs ranging from 5 to 50 mm Hg, and a linear correlation was demonstrated between the two methods (r2 = 0.7752, p < 0.001; average bias -0.019 ml/mm Hg). The Spiegelberg air-pouch ICP/compliance monitor provides ICP and compliance data that are very similar to those obtained using both gold-standard methods and an intraparenchymal ICP monitor over a range of pathophysiological ICPs. The automated closed Spiegelberg system offers practical advantages for the measurement of intraventricular compliance. Assessment of the clinical utility and robustness of the Spiegelberg system, together with the development of an intraparenchymal device, would enhance the clinical utility of automated compliance measurement and expand the range of its applications.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Laboratory testing of the Spiegelberg brain pressure monitor: a technical report.

OBJECTIVES The Spiegelberg brain pressure monitor is a low cost intracranial pressure monitoring system that has been used clinically for some time, mainly in Germany. To provide a rigorous bench comparison of the Spiegelberg monitor with the Camino pressure monitor an evaluation programme has been carried out in the UK Shunt Evaluation Laboratory. DESIGN Drift over 72 hours and with temperat...

متن کامل

A Colombian-based survey on ventriculostomy and intracranial pressure monitor placement practices

Objective: Monitoring of intracranial pressure (ICP) is a routine procedure in the neurosurgery field. Although the routine practice of monitoring ICP has decreased in recent years, the practice patterns have not been studied in Colombia. This study was designed to evaluate the current practice forventriculostomy and ICP monitoring by neurosurgeons in Colombia. Methods...

متن کامل

تغییرات فشار داخل جمجمه ناشی از برخی فعالیت‌های بهداشتی – مراقبتی

  Changes of Intracranial Pressure Due to Some Curative and Hygienic Activities   A. Mahmoodi Rad MSc1, M. Foroozi MSc2*, S. Mohamad alizaheh MSc3   1- Academic Member, School of Nursing, Lar, Iran   2- Academic Member, School of Nursing and Midwifery, Kerman, Iran   3- Academic Member, School of Nursing and Midwifery, Kerman, Iran     Background: Evaluation the patients for determining the sin...

متن کامل

Multi-centre assessment of the Spiegelberg compliance monitor: preliminary results.

Acute brain injury states (e.g. head injury, subarachnoid haemorrhage) show clear inverse relationships of ICP vs compliance, with ICP instability at times of lower compliance states. Variance in compliance values is large in hydrocephalus where ICP is relatively lower and compliance higher. Nonetheless, early experience shows that compliance data influence decisions on CSF diversion treatments...

متن کامل

Clinical comparison of the Spiegelberg parenchymal transducer and ventricular fluid pressure.

The Spiegelberg brain pressure catheter is a low cost implantable intracranial pressure measuring system which has the unique ability to perform regular automatic zeroing. A new version of the catheter has become available with a subdural bolt fixation to allow insertion of the device into the brain parenchyma. The accuracy of this system has been evaluated in comparison with a ventricular flui...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of neurosurgery

دوره 93 6  شماره 

صفحات  -

تاریخ انتشار 2000