Evaluation of automated enzyme immunoassays for five anticonvulsants and theophylline adapted to a centrifugal analyzer.

نویسندگان

  • N Urquhart
  • W Godolphin
  • D J Campbell
چکیده

We report a clinical evaluation of the enzyme immunoassay (EMIT) performed with the GEMSAEC centrifugal analyzer as compared to gas-liquid and liquid chromatography for anticonvulsant drugs and theophylline, respectively. A good correlation was obtained for all drugs, although some difficulties were experienced with one lot of reagent for ethosuximide. The analyzer has an economic advantage if many samples are being analyzed for few drugs in each sample.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Homogeneous enzyme immunoassay for theophylline with a centrifugal analyzer.

A rapid homogeneous enzyme immunoassay for theophylline has been adapted to a centrifugal analyzer. The sample volume used was 10 microliter and sera from as many as 28 patients could be measured in 180 seconds. The assay temperature was 30 degrees and the absorbance rates were measured at 340 nm. Within-run and run-to-run precision showed a coefficient of variation of less than 2.6 and 7.5 per...

متن کامل

Adaptation of EMIT drug assays to a random-access automated clinical analyzer.

Enzyme multiplied immunoassay technique (EMIT) assays for theophylline, carbamazepine, phenytoin, phenobarbital, valproic acid, acetaminophen, gentamicin and tobramycin, have been adapted to the Technicon RA-1000 random access analyzer. Working reagents are stable for a period of at least one week, and calibration need only be performed when preparing fresh reagents. Between-day precision range...

متن کامل

Quantitation of haptens by homogeneous immunoprecipitation. 2. Centrifugal analysis for tobramycin, phenobarbital, and theophylline in serum.

Our homogeneous immunoprecipitation inhibition assay (Clin. Chem. 28:659-661, 1982) is applied here to tobramycin, phenobarbital, and theophylline. Only 10-50 microL of test sample is needed. No sample treatment, dilution, or extraction is required. A test serum sample is simultaneously mixed with a drug conjugate and its specific antiserum in a centrifugal analyzer. The subsequent reaction and...

متن کامل

Performance characteristics of three automated immunoassays for thyroid hormones

Background: Since the introduction of the first radioimmunoassay, several improvements have been made in the design of immunoassays such as method of antibody production, labeling, automation and detection technology. We performed an analytical evaluation of the new electrochemiluminescent immunoassay (ECLIA) for serum TSH, FT4 and T3 in the Elecsys 2010 immunoassay system and compared the resu...

متن کامل

Automated fluorometer/photometer system for homogeneous immunoassays.

A fully automated bench-top clinical analyzer (OPTIMATE TM; Ames/Gilford) performs homogeneous fluorescent immunoassays, colorimetric immunoassays, and determinations of routine blood analytes; drugs, enzymes, metabolites, specific proteins, and hormones in serum. Unique features include a combination fluorescence/absorbance aspirating thermocuvette, a photon-counting fluorometer/photometer, a ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Clinical chemistry

دوره 25 5  شماره 

صفحات  -

تاریخ انتشار 1979