Laboratory evaluation of four different devices for secretion mobilization: Acapella choice, green and blue versus water bottle.

نویسندگان

  • Gabi Mueller
  • Ines Bersch-Porada
  • Sabrina Koch-Borner
  • Anja M Raab
  • Marga Jonker
  • Michael Baumberger
  • Franz Michel
چکیده

BACKGROUND Secretion removal is a key issue in patients with respiratory diseases, and is known to be most effective at vibration frequencies of ∼ 13 Hz and with the greatest amplitudes possible. The Acapella devices and the water bottle are used for secretion removal in daily clinical practice but without detailed knowledge on optimal settings. The aim of this study was to evaluate the 3 different Acapella devices and the water bottle at various settings and flows to determine the optimal devices and settings for effective secretion removal. METHODS Three different Acapella devices were tested at flows of 6, 12, 20, 30, 40, and 50 L/min, and at all 5 settings. The water bottle was filled with 5, 10, or 15 cm of water, and tested at flows of 3, 6, 10, 12, and 20 L/min. For all devices and combinations of settings, we measured the frequency and amplitude of the vibrations, as well as the required pressure to generate vibrations. RESULTS Setting 4 was the best for all 3 Acapella devices, and the filling height of the water bottle should be 5 cm. At these settings, all devices elicited vibration frequencies between 12 and 15 Hz, which is theoretically optimal for secretion mobilization. The resistance pressures of the devices to elicit these vibrations were between 5 and 11 cm H2O. However, the Acapella devices elicit higher vibration amplitudes (5-8 cm H2O) than the water bottle (1.8 cm H2O) CONCLUSIONS:: Setting 4 was optimal for all 3 Acapella devices. The Acapella devices may be more efficient for secretion mobilization than the water bottle, because they elicit greater amplitude of vibrations.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

In vitro evaluation of positive expiratory pressure devices attached to nebulizers.

BACKGROUND Patients with cystic fibrosis perform airway clearance techniques and receive nebulized medications on a regular basis. Some positive expiratory pressure (PEP) devices allow concomitant administration of aerosol. I hypothesized that this practice alters the aerosol characteristics and patient dose. I compared the aerosol characteristics and patient dose of nebulized albuterol from 2 ...

متن کامل

Performance comparison of two oscillating positive expiratory pressure devices: Acapella versus Flutter.

BACKGROUND Oscillatory positive expiratory pressure (PEP) with the Flutter device facilitates secretion removal. In the Flutter a steel ball vibrates inside a cone, causing air flow vibration. A new device, the Acapella, uses a counterweighted plug and magnet to create air flow oscillation. The Acapella comes in 2 models: one for patients with expiratory flow > or = 15 L/min and one for < or = ...

متن کامل

Synthesis of activated carbon/zinc oxide nanocomposite from rapeseed plant waste by green chemistry method, evaluation of its performance in Methylene Blue pigment removal and optimization of influencing parameters using design expert

Background and Objective: Due to the water shortages and the presence of industrial pollutants in water resources, wastewater treatment, especially colored wastewater, is essential. The aim of this study was to treat wastewater containing Methylene Blue dye using activated carbon nanocomposite/zinc oxide nanoparticles (ZnO/AC) obtained from canola oil waste by green method. Materials and Metho...

متن کامل

Evaluation of Virtual Water and Water Footprint of Crop Production (Case study: Qaleganj County)

To address the optimal use of available water resources, a revisit of water resource allocation and management is needed. Water use and water footprint of every product varies in different geographical areas due to various climate, agricultural operations and water efficiencies. Thus, more productive management schemes are to be developed. The goal of this study is to assess the virtual water a...

متن کامل

Evaluation of Protamine Level in Human Sperm Samples Using Chromomycin A3 and Aniline Blue Staining

Background: Current microscopic experimental methods cannot diagnose DNA damages present in spermatozoa .Therefore, some methods are needed to address the abnormality of the genetic material status on the sperm samples. As reported by many investigators aniline blue staining technique has been used for identifying sperm chromatin condensation. Also, chromomycin A3 is used for evaluation of the ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Respiratory care

دوره 59 5  شماره 

صفحات  -

تاریخ انتشار 2014