Radiation dose in balloon dacryocystoplasty: a study using Rando phantoms and thermoluminescent dosimetry.
نویسندگان
چکیده
PURPOSE The radiation dose to the lens of the eye, skin, thyroid and brain of patients who underwent diagnostic and interventional radiological procedures of the lacrimal drainage system have been measured with thermoluminescent dosimeters (TLD-100 and TLD-700) by using an adult male and female Rando phantom. All dose values for one second of fluoroscopic exposure and one frame of digital subtraction dacryocystography (DS-DCG) exposure have been obtained individually in the pastero-anterior (PA) and lateral (LAT) projections. MATERIALS AND METHODS An adult male and female Rando phantom was used instead of the patients. The procedures were performed by using an Advantx AFM C-arm unit coupled with a DX Hiline digital image acquisition and processing system. The 6-inch mode of a triplefield image intensifier (II) was used, with a circular collimation of the same or a slightly smaller size. Two different lithium floride (LiF) thermoluminescent chips were used for absorbed dose measurements: TLD700, approximately 4.5 mm in diameter and 0.9 mm in thickness; TLD100, 3.7 mm square and 0.9 mm in thickness. RESULTS The average values of absorbed doses (lens of the eye, skin, thyroid and brain) measured separately with TLD100 and TLD700 dosimeters after irradiation of the male and female Rando phantom are presented for LAT and PA projections. CONCLUSION This study suggests that useful information for dose determination can be obtained by use of the radiation dose to the lens of the eye, thyroid and brain received during radiological procedure of the LDS for one frame of DS-DCG and one second of fluoroscopy.
منابع مشابه
Dose Measurements of Parotid Glands and Spinal Cord in Conventional Treatment of Nasopharyngeal Carcinoma Using RANDO Phantom and Thermoluminescent Dosimeters
Introduction Radiotherapy is regarded as the first treatment of choice for nasopharyngeal carcinoma. Despite the advantages of radiotherapy, patients may suffer from a wide range of side-effects due to the presence of many sensitive normal tissues in these regions. If the absorbed dose exceeds the tolerance level in parotid glands and the spinal cord, myelopathy, Lhermitte's sign and xerostomia...
متن کاملApplication of GEANT4 radiation transport toolkit to dose calculations in anthropomorphic phantoms.
In this paper, we present a novel implementation of a dose calculation application, based on the GEANT4 Monte Carlo toolkit. Validation studies were performed with an homogeneous water phantom and an Alderson-Rando anthropomorphic phantom both irradiated with high-energy photon beams produced by a clinical linear accelerator. As input, this tool requires computer tomography images for automatic...
متن کاملDose Measurements of Parotid Glands and Spinal Cord in Conventional Treatment of Nasopharyngeal Carcinoma Using RANDO Phantom and Thermoluminescent Dosimeters
Introduction Radiotherapy is regarded as the first treatment of choice for nasopharyngeal carcinoma. Despite the advantages of radiotherapy, patients may suffer from a wide range of side-effects due to the presence of many sensitive normal tissues in these regions. If the absorbed dose exceeds the tolerance level in parotid glands and the spinal cord, myelopathy, Lhermitte's sign and xerostomia...
متن کاملDirect Effective Dose Calculations in Pediatric Fluoroscopy-Guided Abdominal Interventions with Rando-Alderson Phantoms – Optimization of Preset Parameter Settings
OBJECTIVE The aim of the study was to calculate the effective dose during fluoroscopy-guided pediatric interventional procedures of the liver in a phantom model before and after adjustment of preset parameters. METHODS Organ doses were measured in three anthropomorphic Rando-Alderson phantoms representing children at various age and body weight (newborn 3.5kg, toddler 10kg, child 19kg). Colli...
متن کاملDevelopment of Computed Tomography Head and Body Phantom for Organ Dosimetry
Introduction: Quality assurance in Computed tomography (CT) centers in developing countries are largely hindered by the unavailability of CT phantoms. The development of a local CT phantom for the measurement of organ radiation absorbed dose is therefore requisite. Material and Methods: Local CT phantoms were designed to meet the standard criteria of 32 cm diameter for body, 16 cm diameter for...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Diagnostic and interventional radiology
دوره 11 3 شماره
صفحات -
تاریخ انتشار 2005