Comparison of performance of various sphygmomanometers with intra-arterial blood-pressure readings.
نویسندگان
چکیده
Seven types of sphygmomanometer were used in random order on each of nine hypertensive patients and the readings compared with simultaneous intra-arterial blood-pressure recordings. All the devices gave significantly different values for systolic pressure, and only two measured diastolic pressure without significant error. Systolic pressure was consistently underestimated (range 31-7 mm Hg), and all but one instrument overestimated diastolic pressure (range 10-2 mm Hg). The variability of readings was least with the standard mercury sphygmomanometer and the random-zero machine, while with some of the more automated devices single readings were in error up to -68/33 mm Hg. The strong correlations found between intra-arterial and cuff systolic pressures with all devices tested and significant correlations for diastolic pressure with all but one device indicate that, with one possible exception, the sphygmomanometers would give accurate results where a change in blood pressure was the main concern.
منابع مشابه
Comparison of Various Blood Pressure Measurement Techniques with Standard Office Based Blood Pressure Mesurement
Introduction: In order to control Hypertension, it is necessary to understand the methods of blood pressure measurement. Recent guidelines emphasize on employing appropriate blood pressure measurement techniques and white coat hypertension. Definition and methods of blood pressure measurement have changed in the American guideline based on SPRINT trial (Systolic Blood Pressure Intervention Tria...
متن کاملArterial stiffness as underlying mechanism of disagreement between an oscillometric blood pressure monitor and a sphygmomanometer.
Oscillometric blood pressure devices tend to overestimate systolic blood pressure and underestimate diastolic blood pressure compared with sphygmomanometers. Recent studies indicate that discrepancies in performance between these devices may differ between healthy and diabetic subjects. Arterial stiffness in diabetics could be the underlying factor explaining these differences. We studied diffe...
متن کاملEvaluation of noninvasive blood pressure monitoring devices Spacelabs 90202 and 90207 versus resting and ambulatory 24-hour intra-arterial blood pressure.
This study evaluated the accuracy of blood pressure values provided by the Spacelabs 90202 and 90207 devices in comparison with intra-arterial recording in 19 subjects at rest and in nine subjects in ambulatory conditions (Oxford method). At rest Spacelabs monitors reflected intra-arterial systolic blood pressure values very closely but overestimated to a considerable extent intra-arterial dias...
متن کاملAmbulatory blood pressure monitoring.
IN this issue Pickering and Harshfield review the use of invasive and noninvasive arterial blood pressure monitoring in ambulant humans. Their review is well balanced and comprehensive but is limited to the discussion of patients with hypertension. Although it is true that the Framingham Study, the Hypertension Detection and Follow-up Program, and the insurance industry data all show the powerf...
متن کاملClinical evaluation of the Infrasonde D4000 blood pressure monitor.
Choosing a blood pressure machine from their ever increasing number is often difficult due to lack of performance data. We have evaluated a new automated non-invasive blood pressure recorder, the Infrasonde D4000. In comparison with the Hawksley random zero mercury sphygmomanometer the Infrasonde showed less variability for both systolic and diastolic readings. However, it was found to record s...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- British medical journal
دوره 2 6131 شماره
صفحات -
تاریخ انتشار 1978