Performance comparison of nebulizer designs: constant-output, breath-enhanced, and dosimetric.
نویسندگان
چکیده
INTRODUCTION Design differences among pneumatically powered, small-volume nebulizers affect drug disposition (percentage of the dose delivered to the patient, lost to deposition in the equipment, and lost via exhalation to ambient air) and thus affect drug availability and efficacy. OBJECTIVE Evaluate in vitro the dose disposition with 5 nebulizer models, of 3 types (constant-output, breath-enhanced, and dosimetric), using simulated normal, adult breathing. METHODS We compared 5 nebulizer models: 2 constant-output (Misty-Neb and SideStream), 1 breath-enhanced (Pari LCD), and 2 dosimetric (Circulaire and AeroEclipse). Each nebulizer was filled with a 3-mL unit-dose of albuterol sulfate and powered by oxygen at 8 L/min. The nebulizers were connected to an induction throat, connected to a breathing simulator. We measured (1) inhaled drug (subdivided into mass deposited in the induction throat and mass deposited in the filter at the distal end of the induction throat), (2) exhaled drug (lost to the ambient air), (3) drug lost to deposition in the apparatus, and (4) drug left in the unit-dose bottle. The duration of nebulization (until sputter) was measured with a stopwatch. All drug amounts were analyzed via spectrophotometry and expressed as a percentage of the total dose. RESULTS The mean +/- SD inhaled drug percentages were: Misty-Neb 17.2 +/- 0.4%, SideStream 15.8 +/- 2.8%, Pari LCD 15.2 +/- 4.2%, Circulaire 8.7 +/- 1.0%, and AeroEclipse 38.7 +/- 1.3%. The mean +/- SD percentages of drug lost to the ambient air were: Misty-Neb 26.8 +/- 0.7%, SideStream 17.3 +/- 0.4%, Pari LCD 18.3 +/- 0.8%, Circulaire 12.3 +/- 0.8%, and AeroEclipse 6.6 +/- 3.3%. The mean +/- SD percentages of drug lost to deposition in the apparatus were: Misty-Neb 52.3 +/- 0.6%, SideStream 63.4 +/- 3.0%, Pari LCD 62.5 +/- 4.0%, Circulaire 75.8 +/- 0.5%, and AeroEclipse 51.0 +/- 2.1%. Duration of nebulization was shortest with the Circulaire and longest with the AeroEclipse (p < 0.05 via 1-way analysis of variance). CONCLUSIONS The nebulizers we tested differ significantly in overall drug disposition. The dosimetric AeroEclipse provided the largest inhaled drug mass and the lowest loss to ambient air, with the test conditions we used. method.
منابع مشابه
Influence of nebulizer type with different pediatric aerosol masks on drug deposition in a model of a spontaneously breathing small child.
BACKGROUND The performance of nebulizers varies with the design type as well as the breathing patterns of various age groups. The present study quantified aerosol delivery using spontaneously breathing parameters of a small child (2-4 years) by a lung simulator to determine the influence of nebulizer type, actuation mechanisms, and pediatric aerosol masks. METHODS Three types of nebulizer (c...
متن کاملDrug output from nebulizers is dependent on the method of measurement.
The objective of this study was to determine whether current regulatory methods for assessing the output of nebulizers are appropriate for the delivery of nebulized steroid suspensions to patients. We studied a conventional jet nebulizer (the Intersurgical Cirrus), an open-vent nebulizer (the Medicaid Sidestream) and a breath-enhanced nebulizer (the Pari LC Plus), using a constant sampling flow...
متن کاملIn vitro assessment of inter and intra batch variability of a breath-enhanced jet nebulizer
Background Guidelines recommend that patients presenting with acute severe asthma with life-threatening features receive the necessary high doses of b2-agonists via the nebulized route.The SideStream Plus (SS+; Philips Respironics) is a breath-enhanced nebulizer designed to maximize respirable output, while minimizing treatment time [1]. Prior to the introduction of a new version of nebulizer, ...
متن کاملIs the measurement of delivered dose a good performance measure? A characterization of delivered and respirable delivered dose in two brands of jet nebulizer
Background Use of a nebulizer facilitates rapid administration of high doses of drug, which is desirable for patients presenting with acute severe asthma. Breath-enhanced jet nebulizers direct the patient’s air flow through the nebulizer handset during inhalation producing aerosol output with a more favorable respirable profile. SideStream Plus (SS+; Philips Respironics) and Nebutech (NB; Salte...
متن کاملBreath-actuated nebulizer for patients with exacerbation of COPD: efficiency and cost.
Aerosolized bronchodilators have been the first line therapy for exacerbation of COPD (ECOPD) to relieve bronchoconstriction and shortness of breath. Administration of bronchodilators in acute settings is commonly done with small-volume nebulizers, which deliver higher inhaled drug mass, compared to other devices such as pressurized metered-dose inhaler (pMDI) and dry powder inhaler (DPI).1 Sma...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Respiratory care
دوره 49 2 شماره
صفحات -
تاریخ انتشار 2004