Incidence of Deformation and Fracture of Twisted File Adaptive Instruments after Repeated Clinical Use
نویسندگان
چکیده
OBJECTIVES The aim of the present study was to investigate the incidence of deformation and fracture of twisted file adaptive nickel-titanium instruments after repeated clinical use and to identify and check whether the three instruments within the small/medium sequence showed similar or different visible signs of metal fatigue. MATERIAL AND METHODS One-hundred twenty twisted file adaptive (TFA) packs were collected after clinically used to prepare three molars and were inspected for deformations and fracture. RESULTS The overall incidence of deformation was 22.2%, which was not evenly distributed within the instruments: 15% for small/medium (SM)1 (n = 18), 38.33% for SM2 (n = 46) and 13.33% for the SM3 instruments (n = 16). The defect rate of SM2 instruments was statistically higher than the other two (P < 0.001). The fracture rate was 0.83% (n = 3), being two SM2 instruments and one SM3. CONCLUSIONS It was observed a very low defect rate after clinical use of twisted file adaptive rotary instruments. The untwisting of flutes was significantly more frequent than fracture, which might act as prevention for breakage. The results highlight the fact that clinicians should be aware that instruments within a sequence might be differently subjected to intracanal stress.
منابع مشابه
Structural analysis of cyclic-loaded nickel-titanium rotary instruments by using resonance frequency as a parameter.
INTRODUCTION The aim of this study was to investigate the relationship between fatigue life and resonance frequency (RF) of various types of nickel-titanium (NiTi) rotary instruments. In addition, the influence of NiTi instruments with different manufacturing methods on cyclic loading was evaluated by using RF as a parameter. METHODS Twenty-eight ProFile instruments and 10 Twisted File instru...
متن کاملEx Vivo Comparison of File Fracture and File Deformation in Canals with Moderate Curvature: Neolix Rotary System versus Manual K-files
Background and Aim: Cleaning and shaping is one of the important steps in endodontic treatment, which has an important role in root canal treatment outcome. This study evaluated the rate of file fracture and file deformation in Neolix rotary system and K-files in shaping of the mesiobuccal canal of maxillary first molars with moderate curvature. Materials and Methods: In this ex vivo exp...
متن کاملComparative Evaluation of the Effect of Manufacturing Process on Distortion of Rotary ProFile and Twisted File: An in Vitro SEM Study
Background and aims. The manufacturing process of rotary Ni-Ti file can influence its resistance to fracture. The rotary ProFile (Dentsply-Maillefer, Baillagues, Switzerland) is manufactured by grinding mechanism whereas Twisted File (Sy-bron Endo, USA) is manufactured with a twisting method. The purpose of this study was to comparatively evaluate the effect of manufacturing process on distorti...
متن کاملDeformation and fracture incidence of Reciproc instruments: a clinical evaluation.
AIM To investigate the incidence of fracture and deformation observed on Reciproc instruments after clinical use. METHODOLOGY A total of 1696 Reciproc NiTi instruments that were discarded after normal clinical usage over 30 months were collected. Of these 1580 instruments were R25 (93%), 76 were R40 (5%), and 40 were R50 (2%). The treatments included 3780 root canal preparations, 3023 primary...
متن کاملTransportation Assessment in Simulated Curved Canals after preparation with Twisted File Adaptive and BT-Race instruments
BACKGROUND This study compared the incidence of deviation along curved canals after preparation with two nickel-titanium (NiTi) rotary systems, Twisted File Adaptive and BT-RaCe. MATERIAL AND METHODS Forty resin training blocks with curved canals were filled with ink and divided into two groups according to the instrumentation technique. Preinstrumentation images were acquired by using a ster...
متن کامل