Poor results of the Freeman uncemented metal-backed acetabular component. Five-to-nine-year results.

نویسندگان

  • S F Journeaux
  • D A Morgan
  • W J Donnelly
چکیده

We reviewed 120 consecutive primary total hip replacements in 109 patients in whom a Freeman uncemented metal-backed component had been used. Acetabular components were used with a Freeman neck-retaining stem in one of three configurations: cemented smooth stem, uncemented smooth stem or uncemented ridged stem. After a mean follow up of 72 months (62 to 113) there were nine cases of aseptic loosening; a survivorship at eight years of 83%. In addition, more than one-third of the remaining surviving cups showed loosening radiologically. The Freeman acetabular component demonstrated a characteristic pattern of loosening on radiographs. The high incidence of aseptic loosening can be explained by poor design and material failure.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Failure of an uncemented non-porous metal-backed prosthesis with augmentation using impacted allograft for acetabular revision 12- to 17-year results.

We have reviewed 29 patients (30 hips) who had undergone revision total hip arthroplasty using a Freeman metal-backed acetabular component and acetabular impaction allografting. The mean follow-up was for 153 years (12 to 17). Five patients (5 hips) died with the prosthesis in situ and four (4 hips) were lost to follow-up. Twelve hips had failed and in the remaining nine there were minor sympto...

متن کامل

Metal-backed acetabular components with conventional polyethylene: a review of 9113 primary components with a follow-up of 20 years.

The Norwegian Arthroplasty Register has shown that several designs of uncemented femoral stems give good or excellent survivorship. The overall findings for uncemented total hip replacement however, have been disappointing because of poor results with the use of metal-backed acetabular components. In this study, we exclusively investigated the medium-to long-term performance of primary uncement...

متن کامل

Polyethylene wear in uncemented acetabular components.

We measured polyethylene wear in 231 porous-coated uncemented acetabular cups. We divided the hips into two groups according to the fixation of the femoral component, by cementing (n = 97) or press-fit (n = 134). Follow-up was from three to five years. The patients in two sub-groups were matched for weight, diagnosis, sex, age and length of follow-up. The linear wear rate of cups articulated wi...

متن کامل

Polyethylene wear in cemented metal-backed acetabular cups.

We examined radiographic polyethylene wear in 233 cemented total hip arthroplasties (201 patients) with either a metal-backed or a non-metal-backed acetabular cup. All patients had identical cemented one-piece titanium femoral stems with a femoral head diameter of 28 mm. The mean linear wear rate was 0.11 mm/yr in metal-backed sockets and 0.08 mm/yr in non-metal-backed sockets (p = 0.0002). The...

متن کامل

Polyethylene wear in total hip arthroplasty.

Polyethylene wear increasingly has been blamed for osteolysis and granuloma formation after total joint arthroplasty. The authors evaluated the wear difference between bearing surfaces of stainless steel, cobalt chrome, and titanium alloy. They also compared cemented all-polyethylene molded cups with cemented metal-backed molded acetabular cups. These were compared with molded and machined poly...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The Journal of bone and joint surgery. British volume

دوره 82 2  شماره 

صفحات  -

تاریخ انتشار 2000