Sealing of Vessels Larger Than 7 Millimeters Using Enseal in Porcine Aorta

نویسندگان

  • Shahida Bibi
  • Saeed Alblawi
  • Vamsi Velchuru
  • Gopichand Thripuraneni
  • Francisco Quinteros
  • Jasna Coralic
  • Slawomir Marecik
  • John Park
  • Leela M. Prasad
چکیده

BACKGROUND The Enseal (Ethicon Endo-Surgery, Blue Ash, Ohio) tissue-sealing device has proven efficacy for ligation of vessels<7 mm in diameter, even with significant supraphysiologic bursting pressures. We aimed to evaluate the safety of Enseal in porcine vessels>7 mm. MATERIALS AND METHODS The lumbar aortas of pigs that were euthanized for unrelated procedures were harvested. A 5- to 6-cm segment of aorta was sealed using the Enseal device. The opposite end was attached to a pressure-testing device to measure pressures at leak or bursting. The bivariate Pearson correlation was used to determine the relationship between diameter and bursting pressure. One-way analysis of variance was used to determine differences between the groups of vessels on the basis of their diameter. RESULTS Ninety samples of 5-cm aorta segments were used to assess bursting pressure. The median diameter was 14 mm (range, 7-18) and bursting pressure was 85 mm Hg (range, 24-650). The Pearson test showed a negative correlation between vessel diameter and bursting pressure (P=.25). One-way analysis of variance did not show any significant difference between vessel diameters grouped by size (P=.517), and neither did the Scheffe post hoc test when comparing diameter with bursting pressure; 31% of specimens failed to seal. CONCLUSIONS Bursting pressures are low and inconsistent after tissue sealing with the Enseal device in porcine vessels>7 mm. These vessels also demonstrated a higher rate of failure to seal. The histologic results of the aorta segments (ie, a low collagen-elastin ratio) may be the cause of the low bursting pressures.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Consistency and sealing of advanced bipolar tissue sealers

OBJECTIVES The aim of this study was to evaluate two commonly used advanced bipolar devices (ENSEAL(®) G2 Tissue Sealers and LigaSure™ Blunt Tip) for compression uniformity, vessel sealing strength, and consistency in bench-top analyses. METHODS Compression analysis was performed with a foam pad/sensor apparatus inserted between closed jaws of the instruments. Average pressures (psi) were rec...

متن کامل

Laparoscopic Nephrectomy Using the EnSeal Tissue Sealing and Hemostasis System: Successful Therapeutic Application of Nanotechnology

The potential impact of nanotechnology in the field of urology is broad with diagnostic and therapeutic benefits that have only recently begun to be explored. Application of nanotechnology principles to tissue and vessel sealing during laparoscopic procedures may reduce associated thermal injury and inflammatory response. We report our initial experience using the EnSeal Tissue Sealing and Hemo...

متن کامل

Sealing vessels up to 7 mm in diameter solely with ultrasonic technology

INTRODUCTION Ultrasonic energy is a mainstay in the armamentarium of surgeons, providing multifunctionality, precision, and control when dissecting and sealing vessels up to 5 mm in diameter. Historically, the inability to seal vessels in the 5-7 mm range has been perceived as an inherent limitation of ultrasonic technology. The purpose of this study was to evaluate sealing of vessels up to 7 m...

متن کامل

Rapid sealing and cutting of porcine blood vessels, ex vivo, using a high-power, 1470-nm diode laser.

Suture ligation with subsequent cutting of blood vessels to maintain hemostasis during surgery is time consuming and skill intensive. Energy-based electrosurgical and ultrasonic devices are often used to replace sutures and mechanical clips to provide rapid hemostasis and decrease surgery time. Some of these devices may create undesirably large collateral zones of thermal damage and tissue necr...

متن کامل

Rationale, bench testing and in vivo evaluation of a novel 5 mm laparoscopic vessel sealing device with homogeneous pressure distribution in long instrument jaws

BACKGROUND In 1998, an electrothermal bipolar vessel sealing (EBVS) system was introduced and quickly became an integral component of the surgical armamentarium in various surgical specialties. Currently available EBVS instruments use a scissor-like jaw configuration and closing mechanism, which causes decreasing compression pressure from the proximal to the distal end of the jaws. A new EBVS s...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 18  شماره 

صفحات  -

تاریخ انتشار 2014