Accuracy of the i-STAT bedside blood gas analyser.

نویسندگان

  • S Sediame
  • F Zerah-Lancner
  • M P d'Ortho
  • S Adnot
  • A Harf
چکیده

The performance of the i-STAT portable clinical analyser for measuring blood gases and pH was evaluated with reference to a conventional blood gas analyser (ABL520 Radiometer). Ninety-two samples from the routine blood gas analysis laboratory were chosen according to a wide distribution of partial pressure of carbon dioxide (Pa,CO2), partial pressure oxygen (Pa,O2) and pH and then analysed. All measurements were performed in duplicate by trained technicians from the central hospital laboratory. Differences between duplicate measurements were computed for Pa,CO2: (1.2 versus 0.4%), Pa,O2 (1.7 versus 1.1%) and pH (0.06 versus 0.02%), for the i-STAT and ABL520, respectively. pH and Pa,CO2 values measured with the i-STAT were very close to those obtained with the ABL520, the difference (mean+/-SD) being 0.006+/-0.018 and -0.13+/-0.17 kPa, respectively. Statistical analysis showed that the differences between analysers did not depend on values of pH or Pa,CO2. The performance of the analysers depended on the level of PO2. Below 15 kPa (n=48), the two systems gave nearly identical values, the mean difference was 0.01+/-0.37 kPa. Between 16 and 55 kPa (n=44), there was a systematic but small (-0.69+/-0.67 kPa) underestimation of Pa,O2 measured with the i-STAT (p<10(-8)). In conclusion, this study shows that blood gas analysis using the i-STAT portable device is comparable with that performed by a conventional laboratory blood gas analyser.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of the Radiometer ‘ABL4’ pH/blood gas/K+/Hb analyser

The ABL4 analyser is designed for stat determination of blood gas (pCO2, pO2), pH, potassium and haemoglobin (Hb), a panel particularly adapted to the follow-up of liver transplantation. The performance of the analyser in classical tests was evaluated; in addition, K+ measurements were performed in specific conditions (linearity in a non-physiological range, erythrocyte interference). The evalu...

متن کامل

Evaluation of a new instrument offering sodium, potassium, and ionized calcium in combination with haematocrit, pH and blood gas analysis; the performance of the Nova Stat Profile 1

The Nova Stat Profile 1 analyser, a combined sodium, potassium, ionized calcium, haematocrit and conventional blood gas analyser, was evaluated over a four month period. In addition to assessing and demonstrating that the instrument met analytical requirements, an appreciation of the use of ionised calcium (iCa) was made. Prospective costs were characterised and practical problems of iCa measur...

متن کامل

Evaluation of the i-STAT Portable Clinical Analyzer for point-of-care blood testing in the intensive care units of a university children's hospital.

We evaluated the analytical performance of the i-STAT Portable Clinical Analyzer (PCA), a point-of-care testing system consisting of a hand-held analyzer and single-use cartridges that measure different panels of electrolytes, metabolites, blood gases, and hematocrit in 65-100 microl of blood. Our objective was to determine whether PCA measurements at the bedside of patients in the neonatal and...

متن کامل

Validation of the i-STAT system for the analysis of blood gases and acid–base status in juvenile sandbar shark (Carcharhinus plumbeus)

Accurate measurements of blood gases and acid-base status require an array of sophisticated laboratory equipment that is typically not available during field research; such is the case for many studies on the stress physiology, ecology and conservation of elasmobranch fish species. Consequently, researchers have adopted portable clinical analysers that were developed for the analysis of human b...

متن کامل

Analytical performance of the i-STAT cardiac troponin I assay.

BACKGROUND This study determines the analytical characteristics of the i-STAT cardiac troponin I assay (cTnI; i-STAT, Princeton, NJ), a 10-min POC assay, designed to be performed at the bedside. METHODS Three different hospitals participated in a patient specimen and analytical validation study (n=186) for the i-STAT cTnI assay carried out in real time. A total of 186 whole blood specimens (l...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The European respiratory journal

دوره 14 1  شماره 

صفحات  -

تاریخ انتشار 1999