Microenvironment in Ventilated Animal Cages with Differing Ventilation Rates, Mice Populations, and Frequency of Bedding Changes.
نویسندگان
چکیده
The purpose of the study was to assess the microenvironment in separately ventilated mouse cages at differing ventilation rates, mice populations, and frequency of bedding changes. We monitored intracage temperature, relative humidity, and concentrations of ammonia and carbon dioxide during 3 experiments. First, the effect of ventilation rate on the microenvironment of cages housing adult male mice was evaluated at 30, 40, 60, 80 and 100 air changes/h. For all ventilation rates tested, ammonia concentration was less than 3 ppm, carbon dioxide concentration ranged from 840 to 3,300 ppm, relative humidity ranged from 42 to 65%, and temperature ranged from 23.2 to 25.3 C. Second, we monitored cage microenvironment continually in experiments during which changing of bedding was delayed. Male mice were used in the experiment, and cages were ventilated (60 air changes/h). Cages were allowed to accumulate soiled material for 26 days, during which time ammonia concentration and relative humidity did not exceed 10 ppm and 45%, respectively. Third, we tested ventilation rate and frequency of bedding changes in ventilated cages containing breeding trios (2 females, 1 male) and their pups. Ammonia concentrations remained at 25 ppm or less for 30, 60, and 100 air changes/h when bedding was changed weekly and for 100 air changes/h when bedding was changed every 2 weeks. We concluded that 30 air changes/h was sufficient to maintain a healthful microenvironment in cages that were housing adult male mice in which bedding was changed weekly. When frequency of bedding changes was reduced to every 2 weeks, 60 air changes/h was sufficient for cages housing adult males, but 100 air changes/h was necessary for cages housing breeding trios and pups.
منابع مشابه
Comparison of environment and mice in static and mechanically ventilated isolator cages with different air velocities and ventilation designs.
The purpose of this study was to compare environmental conditions and mice in cages with four different mechanical ventilation designs and a static isolator cage. Environmental conditions (air velocity, temperature, relative humidity, bedding weight change, airborne dust, NH3, and CO2) were compared for each cage type (n = 5 per cage). Bedding type was chipped hardwood. Mouse response in each c...
متن کاملEvaluation of cage micro-environment of mice housed on various types of bedding materials.
A variety of environmental factors can affect the outcomes of studies using laboratory rodents. One such factor is bedding. Several new bedding materials and processing methods have been introduced to the market in recent years, but there are few reports of their performance. In the studies reported here, we have assessed the cage micro-environment (in-cage ammonia levels, temperature, and humi...
متن کاملIndividually ventilated cages microclimate monitoring using photoacoustic spectroscopy
A multi-gas photoacoustic sensor was developed in order to monitor microclimatic parameters at ppm level from the atmosphere of individually ventilated cages housing mice. Ammonia, water vapour and carbon dioxide were measured on-line and in real-time during two weeks and 24 hours a day. The quality of bedding and the ventilation rate inside the cages has been monitored. The circadian activity ...
متن کاملEffect of 2 Bedding Materials on Ammonia Levels in Individually Ventilated Cages.
This study sought to identify an optimal rodent bedding and cage-change interval to establish standard procedures for the IVC in our rodent vivarium. Disposable cages were prefilled with either corncob or α-cellulose bedding and were used to house 2 adult Sprague-Dawley rats (experimental condition) or contained no animals (control). Rats were observed and intracage ammonia levels measured dail...
متن کاملMass Generation Rates of Ammonia, Moisture, and Heat Production in Mouse Cages with Two Bedding Types, Two Mouse Strains, and Two Room Relative Humidities
Mass generation rates of ammonia, moisture, and heat production were measured for mice that were housed in standard shoebox cages at a density of five mice per cage and maintained in either a 35% or 75% room relative humidity. Two mouse strains kept on two different types of bedding materials were compared when bedding material had been in the cages for two, six, and ten days. Gas exchanges dir...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Contemporary topics in laboratory animal science
دوره 37 2 شماره
صفحات -
تاریخ انتشار 1998