Dosimetric characterization of Elekta stereotactic cones

نویسندگان

  • Egor Borzov
  • Alexander Nevelsky
  • Raquel Bar‐Deroma
  • Itzhak Orion
چکیده

PURPOSE Dosimetry of small fields defined by stereotactic cones remains a challenging task. In this work, we report the results of commissioning measurements for the new Elekta stereotactic conical collimator system attached to the Elekta VersaHD linac and present the comparison between the measured and Monte Carlo (MC) calculated data for the 6 MV FFF beam. In addition, relative output factor (ROF) dependence on the stereotactic cone aperture variation was studied and penumbra comparison for small MLC-based and cone-based fields was performed. METHODS Cones with nominal diameters of 15 mm, 12.5 mm, 10 mm, 7.5 mm, and 5 mm were employed in our study. Percentage depth dose (PDD), off-axis ratios (OAR), and ROF were measured using a stereotactic field diode (SFD). BEAMnrc code was used for MC simulations. RESULTS MC calculated and measured PDDs for all cones agreed within 1%/0.5 mm, and OAR profiles agreed within 1%/0.5 mm. ROF obtained from the measurements and MC calculations agreed within 2% for all cone sizes. Small-field correction factors for the SFD detector Kfield,3 × 3 (SFD) were derived using MC calculations as a baseline and were found to be 0.982, 0.992, 0.997, 1.015, and 1.017 for the 5, 7.5, 10, 12.5, and 15-mm cones respectively. The difference in ROF was about 10%, 6%, 3.5%, 3%, 2.5%, and 2% for ±0.3 mm variations in 5, 7.5, 10, 12.5, and 15-mm cone aperture respectively. In case of single static field, cone-based collimation produced a sharper penumbra compared to the MLC-based. CONCLUSIONS Accurate MC simulation can be an effective tool for verification of dosimetric measurements of small fields. Due to the very high sensitivity of output factors on the cone diameter, manufacture-related variations in cone size may lead to considerable variations in dosimetric characteristics of stereotactic cones.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Commissioning of small field size radiosurgery cones in a 6-MV flattening filter-free beam.

This study aimed to describe the commissioning of small field size radiosurgery cones in a 6-MV flattening filter free (FFF) beam and report our measured values. Four radiosurgery cones of diameters 5, 10, 12.5, and 15 mm supplied by Elekta Medical were commissioned in a 6-MV FFF beam from an Elekta Versa linear accelerator. The extraction of a reference signal for measuring small fields in sca...

متن کامل

Delivery efficiency of an Elekta linac under gated operation

In this study, we have characterized the efficiency of an Elekta linac in the delivery of gated radiotherapy. We have explored techniques to reduce the beam-on delay and to improve the delivery efficiency, and have investigated the impact of frequent beam interruptions on the dosimetric accuracy of gated deliveries. A newly available gating interface was installed on an Elekta Synergy. Gating s...

متن کامل

Penumbra Measurements and Comparison of In-House and Standard Circular Cones by the Gafchoromic Film, Pinpoint Ion Chamber, and MCNPX Monte Carlo Simulation

Introduction: Penumbra is an important property of the radiation beam to obtain a suitable margin surrounding the target volume. Therefore, the precise penumbra width determination in stereotactic radiotherapy is necessary for treatment planning. This study aimed to compare the obtained results of penumbra width by in-house and standard circul...

متن کامل

Dosimetric characteristics of the small diameter BrainLab™ cones used for stereotactic radiosurgery

The purpose was to study the dosimetric characteristics of the small diameter (≤ 10.0 mm) BrainLAB cones used for stereotactic radiosurgery (SRS) treatments in conjunction with a Varian Trilogy accelerator. Required accuracy and precision in dose delivery during SRS can be achieved only when the geometric and dosimetric characteristics of the small radiation fields is completely understood. Alt...

متن کامل

Comparison of measured and Monte Carlo calculated dose distributions from “circular collimators” for radiosurgical beams

Background: Stereotactic radiosurgery is an important clinical tool for the treatment of small lesions in the brain, including benign conditions, malignant and localized metastatic tumors. A dosimetry study was performed for Elekta ‘Synergy S ’ as a dedicated Stereotactic radiosurgery unit, capable of generating circular radiation fields with diameters of 1-5 cm at isocentre using the BEAM/EGS4...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 19  شماره 

صفحات  -

تاریخ انتشار 2018