Outcomes for selected patients with spinal metastases may be improved by dose escalation using stereotactic body radiotherapy (SBRT). As target geometry is complex, we compared SBRT plans using step-and-shoot intensity-modulated radiotherapy (IMRT), carbon ion RT, and proton RT. We prepared plans treating cervical, thoracic, and lumbar metastases for three different techniques--IMRT, carbon ion, and proton plans--to deliver a median single 24 Gy fraction such that at least 90% of the planning target volume (PTV) received more than 18 Gy and were compared for PTV coverage, normal organ sparing, and estimated delivery time. PTV coverage did not show significant differences for the techniques, spinal cord dose sparing was lowered with the particle techniques. For the cervical lesion spinal cord maximum dose, dose of 1% (D1), and percent volume receiving 10 Gy (V10Gy) were 11.9 Gy, 9.1 Gy, and 0.5% in IMRT. This could be lowered to 4.3 Gy, 2.5 Gy, and 0% in carbon ion planning and to 8.1 Gy, 6.1 Gy, and 0% in proton planning. Regarding the thoracic lesion no difference was found for the spinal cord. For the lumbar lesion maximum dose, D1 and percent volume receiving 5Gy (V5Gy) were 13.4 Gy, 8.9 Gy, and 8.9% for IMRT; 1.8 Gy, 0.7 Gy, and 0% for carbon ions; and 0 Gy, < 0.01 Gy, and 0% for protons. Estimated mean treatment times were shorter in particle techniques (6-7 min vs. 12-14 min with IMRT). This planning study indicates that carbon ion and proton RT can deliver high-quality PTV coverage for complex treatment volumes that surround the spinal cord.
منابع مشابه
Radiation-induced toxicity after image-guided and intensity-modulated radiotherapy versus external beam radiotherapy for patients with spinal bone metastases (IRON-1): a study protocol for a randomized controlled pilot trial
BACKGROUND Radiation therapy (RT) of bone metastases provides an important treatment approach in palliative care treatment concepts. As a consequence of treatment, the extent of radiation-induced toxicity is a crucial feature with consequences to a patient's quality of life. In this context this study aims at reducing the extent of radiation-induced side effects and toxicity by assuming a bette...
متن کامل3-Dimensional conformal radiotherapy versus intensity modulated radiotherapy for localized prostate cancer: Dosimetric and radiobiologic analysis
Background: To analyze the dosimetric and radio biologic advantages between intensity modulated radiotherapy (IMRT) and 3 dimensional conformal radiotherapy (3DCRT) and selection of optimal photon energy for IMRT treatments. Material and methods: 24 patients with localized prostate carcinoma were planned for 3DCRT and IMRT techniques. Radiation dose of 54 Gy with 2 Gy/fraction, was planned to ...
متن کاملDosimetric Evaluation of Volumetric Modulated Arc Therapy (VMAT) and Intensity Modulated Radiotherapy (IMRT) Using AAPM TG 119 Protocol
Background: The commissioning accuracy of Volumetric Modulated Arc Therapy (VMAT) need to be evaluated.Objective: To test and evaluate commissioning accuracy of VMAT based on the TG 119 protocols at local institution. Material and Methods: The phantom, structure sets, VMAT and IMRT beam parameter setup, dose prescriptions and planning objectives were following TG 119 guidelines to c...
متن کاملComparison of intensity modulated radiotherapy (IMRT) with intensity modulated particle therapy (IMPT) using fixed beams or an ion gantry for the treatment of patients with skull base meningiomas
BACKGROUND To examine the potential improvement in treatment planning for patients with skull base meningioma using IMRT compared to carbon ion or proton beams with and without a gantry. METHODS Five patients originally treated with photon IMRT were selected for the study. Ion beams were chosen using a horizontal beam or an ion gantry. Intensity controlled raster scanning and the intensity mo...
متن کاملStep-and-Shoot versus Compensator-based IMRT: Calculation and Comparison of Integral Dose in Non-tumoral and Target Organs in Prostate Cancer
Introduction Intensity-Modulated Radiotherapy (IMRT) is becoming an increasingly routine treatment method. IMRT can be delivered by use of conventional Multileaf Collimators (MLCs) and/or physical compensators. One of the most important factors in selecting an appropriate IMRT technique is integral dose. Integral dose is equal to the mean energy deposited in the total irradiated volume of the p...