Determination of Total Hip Replacement stem insertion endpoint and stability assessment by vibration analysis: first experiences with per-operative measurements

نویسندگان

  • Leonard C. Pastrav
  • Siegfried V. N. Jaecques
  • Michiel Mulier
  • Georges Van der Perre
چکیده

The assessment of the primary stability of cementless implants still remains a subjective factor in total hip replacement (THR) and, as a consequence, the excessive press-fitting of a THR component can be the cause of intra-operative fractures. Objective information about the stability of implant-bone structures can be obtained using methods based on vibration analysis. After extensive in vitro studies, a per-operative protocol was designed to detect the insertion endpoint and/or to assess the stability of custom made hip prostheses. The experiments were performed on cementless hip stems and on hybrid hip stems that were partially cemented distally. This paper presents the frequency response function evolution during the hip stem insertion in the femur, in per-operative conditions.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Per-operative vibration analysis: a valuable tool for defining correct stem insertion: preliminary report.

BACKGROUND Defining the stem insertion end point during total hip replacement still relies on the surgeon's feeling. When a custom-made stem prosthesis with an optimal fit into the femoral canal is used, the risk of per-operative fractures is even greater than with standard prostheses. Vibration analysis is used in other clinical settings and has been tested as a means to detect optimal stem in...

متن کامل

In vivo evaluation of a vibration analysis technique for the per-operative monitoring of the fixation of hip prostheses

BACKGROUND The per-operative assessment of primary stem stability may help to improve the performance of total hip replacement. Vibration analysis methods have been successfully used to assess dental implant stability, to monitor fracture healing and to measure bone mechanical properties. The objective of the present study was to evaluate in vivo a vibration analysis-based endpoint criterion fo...

متن کامل

Intra-operative monitoring of orthopaedic implant stability by vibration analysis

The success rate of orthopaedic implants is strongly dependent on the initial fixation, but the objective intra-operative assessment of primary stability of cementless implants is still a challenge and, moreover, excessive press-fitting of the implant can cause intraoperative fractures. In a series of in vitro and in vivo experiments the change in the frequency response function (FRF) of the hi...

متن کامل

Total Hip Replacement Revision in a Single Brand Small Cementless Stem – Our Experience after the Findings of the National Joint Registry

Background: Cementless total hip replacement is the common THR performed in England, Wales, Northern Irelandand the Isle of Man. The Corail stem is the most popular cementless implant and has a ODEP 10A rating. Review ofits performance in the registry identified an increase rate of revision amongst the smaller stem sizes. However, claritywas not provided on the explanation for this finding. We ...

متن کامل

Analysis of the fixation quality of cementless hip prostheses using a vibrational technique

Vibration analysis can be used as a more sensitive alternative to conventional techniques for the detection of loosening of total hip replacement (THR) components. The vibration modes of the human femur and of selected THR prosthetic stems were studied by experimental modal analysis using a CADA-X system and by finite element simulation using MARC/Mentat software. Two hypotheses were formulated...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2006