Peripheral snap-fit locking mechanisms and smooth surface finish of tibial trays reduce backside wear in fixed-bearing total knee arthroplasty
نویسندگان
چکیده
Background and purpose - Severe backside wear, observed in older generations of total knee replacements (TKRs), led to redesign of locking mechanisms to reduce micromotions between tibial tray and inlay. Since little is known about whether this effectively reduces backside wear in modern designs, we examined backside damage in retrievals of various contemporary fixed-bearing TKRs. Patients and methods - A consecutive series of 102 inlays with a peripheral (Stryker Triathlon, Stryker Scorpio, DePuy PFC Sigma, Aesculap Search Evolution) or dovetail locking mechanism (Zimmer NexGen, Smith and Nephew Genesis II) was examined. Articular and backside surface damage was evaluated using the semiquantitative Hood scale. Inlays were examined using scanning electron microscopy (SEM) to determine backside wear mechanisms. Results - Mean Hood scores for articular (A) and backside (B) surfaces were similar in most implants-Triathlon (A: 46, B: 22), Genesis II (A: 55, B: 24), Scorpio (A: 57, B: 24), PFC (A: 52, B: 20); Search (A: 56, B: 24)-except the NexGen knee (A: 57, B: 60), which had statistically significantly higher backside wear scores. SEM studies showed backside damage caused by abrasion related to micromotion in designs with dovetail locking mechanisms, especially in the unpolished NexGen trays. In implants with peripheral liner locking mechanism, there were no signs of micromotion or abrasion. Instead, "tray transfer" of polyethylene and flattening of machining was observed. Interpretation - Although this retrieval study may not represent well-functioning TKRs, we found that a smooth surface finish and a peripheral locking mechanism reduce backside wear in vivo, but further studies are required to determine whether this actually leads to reduced osteolysis and lower failure rates.
منابع مشابه
Micromotion at the tibial plateau in primary and revision total knee arthroplasty: fixed versus rotating platform designs.
OBJECTIVES Initial stability of tibial trays is crucial for long-term success of total knee arthroplasty (TKA) in both primary and revision settings. Rotating platform (RP) designs reduce torque transfer at the tibiofemoral interface. We asked if this reduced torque transfer in RP designs resulted in subsequently reduced micromotion at the cemented fixation interface between the prosthesis comp...
متن کاملPolyethylene Backside Damage and Insert Micromotion in Total Knee Arthroplasty:does the Locking Mechanism Degrade with Time?
Introduction: It has been suggested that the capture mechanism of modular polyethylene tibial inserts degrades with time in-situ.[2] However, a significant correlation between insert micromotion and insitu time has not been reported. While it is intuitive that greater micromotion would result in greater backside damage, there are few data from autopsy retrievals to support this point and large ...
متن کاملPolished Cobalt-Chrome vs Titanium Tibial Trays in Total Knee Replacement (a Comparison using the PFC Sigma System)
Fixed-bearing total knee arthroplasty components can cause wear debris due to fretting micromotion between the polyethylene insert and the metal tibial tray, possibly leading to osteolysis and implant failure. This study compared the effects of either a highly polished cobalt-chrome (CoCr) or titanium tibial tray in patients receiving the PFC. Sigma® posterior stabilized knee system with a mode...
متن کاملEffect of Polyethylene Crosslinking and Bearing Design on Wear of Unicompartmental Arthroplasty.
Wear and polyethylene damage continue to be important factors affecting outcomes of unicompartmental knee arthroplasty. We compared two design rationales for unicompartmental arthroplasty: fully congruent mobile bearings; or moderately conforming fixed bearings using experimental and computational wear simulation. Experimental wear rates were 3.89 (±0.12) mg/million cycles for the highly crossl...
متن کاملPatient-Specific Unicompartmental Knee Resurfacing Arthroplasty: Use of a Novel Interference Lock to Reduce Tibial Insert Micromotion and Backside Wear
Micromotion has long been associated with wear of polyethylene tibial inserts, potentially causing failure of unicompartmental knee replacement systems. One cause of micromotion is the locking mechanism between the undersurface of the polyethylene and the tibial tray. The objective of this study was to investigate the use of new novel lock designs for reducing the micromotion associated with a ...