Effects of a Novel Lumbar Non-Fusion Stabilization Device on Adjacent Segment: in vitro & FEM study
ثبت نشده
چکیده
in vitro & FEM study Kiapour A.; Parikh R.; Koruprolu S.; Mhatre D.; +Goel VK; Krishna M.; Friesem T. +Engineering Center for Orthopaedic Research Excellence (ECORE), Departments of Bioengineering and Orthopedics Surgery, Colleges of Engineering and Medicine. University of Toledo, Toledo, OH The University Hospital of North Tees, Stockton-on-Tees, UK Senior author [email protected] INTRODUCTION: Lumbar spinal stenosis is a progressive degenerative condition due to arthritic facet joints. Arthritic facets become inflamed and often develop osteophytes, leading to nerve compression and persistent severe back pain. When conservative treatment fails to reduce pain, surgical management may be pursued to improve the patient's quality of life. Spinal decompression and fusion is one of the most common surgical procedures for treatment of spinal stenosis. However, fusion may result in accelerated degeneration of the adjacent motion segments and morbidity. Motion preservation instrumentation is being developed to preserve motion at the involved and adjacent segments, as opposed to fusion procedure. In this study, we used experimental and finite element (FE) techniques to assess and compare the biomechanics of intact and spines implanted with a novel posterior dynamic stabilizer device (PDS). The effects on the adjacent segment, including motion and intra-discal pressure, were analyzed. METHODS: A cadaveric experiment using six fresh-frozen, healthy L1–S1 ligamentous spine segments was conducted to analyze spine biomechanics under various conditions. The specimens were thawed, and the tissue surrounding the spine, except the ligaments and disc, were carefully removed. The specimens were radiographed for any anomalies or fractures. Each spine was potted using Bondo (a 2-part epoxy resin) cranially at L1 and caudally at S1 (Fig. 1). The base was bolted to a test frame located in the field of view of an Optotrak Motion Measuring System (Northern Digital, Waterloo, Ontario, Canada). The vertebral bodies were affixed with a set of three LEDs each for recording the spatial locations in response to the applied load using the Optotrak system. The IDP sensors were placed at various places within the nucleus of implanted and adjacent segment to record the changes in the nucleus pressure (IDP) during testing. Each specimen was tested in intact and following PDS implant. A pair of PDS was placed at the L4-L5 level using pedicle screws. The PDS device included a set of curved male-female parts with restricted sliding motion imposed by stoppers attached to springs inside the female part. For each condition the specimens were subjected to pure moments to simulate flexion (Flex), extension (Ext), right and left lateral bending (RB, LB), and right and left axial rotation (RR, LR). The applied moment was incrementally increased in 2.5 Nm steps to 10 Nm. The angular motion and IDP at the implanted and adjacent levels were recorded and compared in each case. The Flexion-Extension modes were repeated with same loading configurations plus a follower pre-load of 400 N applied through a system of pulleys and cables. A 3D experimentally validated nonlinear FE model of L3/S1 spine [1] was used to simulate intact and PDS implanted conditions. In order to simulate the implant, a 3D model of PDS was meshed and placed at L4L5 level of intact model. The Cobalt-Chrome material properties were used to define PDS and a frictionless contact pattern was defined between articulation/sliding surfaces within the implant assembly. The implanted and intact models were loaded with 400 N of follower load and 10 Nm of moment in various loading modes. FE model predictions were compared with cadaver data and then used to predict other parameters like maximum stress across the pedicle screws. These were compared with stresses in screws of an existing single level fusion model with similar loading configurations. RESULTS: In the in vitro study, placement of the PDS didn’t alter the motion of the adjacent segment (P>0.5 in all loading cases). The FE model had similar prediction with motions of 4.9, 2.3, 4.7 and 2.6 in Flex, Ext, RB/LB and LR/RR respectively which were close to average and within 1 SD of the in vitro data (Fig.2A). At the index level, placement of implant decreased the motion in all specimens for all loading cases (P<0.02). The motions in the FE model (within 1 SD of cadaveric data) were 5.1, 2.9, 4.5 and 2.6 in intact versus 2.4, 1.2, 1 and 1.1 in PDS model for Flex, Ext, RB/LB and LR/RR respectively. The IDP followed the observed/predicted changes in motion. The IDP values did not have significant changes at the adjacent segment in the implanted conditions for both FE and cadaver experiments with P values ranging from 0.36 to 0.9 for all loading cases in cadaveric experiment. However, there was a significant reduction (P values<0.03), except in Ext (P=0.07) in the IDP compared to intact conditions at the implanted segment. In FE models there were no notable IDP changes for the implanted segment, compared to the intact. The stresses at pedicle screws at both L4 and L5 levels were less in PDS model (L4: 69-127MPa, L5: 67-77MPa) versus fusion (L4: 102163MPa, L5: 120-100MPa) in all loading cases (Fig.2B).
منابع مشابه
Evaluation of a Hybrid Dynamic Stabilization and Fusion System in the Lumbar Spine: A 10 Year Experience
INTRODUCTION The development of adjacent-segment disease is a recognized consequence of lumbar fusion surgery. Posterior dynamic stabilization, or motion preservation, techniques have been developed which theoretically decrease stress on adjacent segments following fusion. This study presents the experience of using a hybrid dynamic stabilization and fusion construct for degenerative lumbar spi...
متن کاملBiomechanical comparison of an interspinous device and a rigid stabilization on lumbar adjacent segment range of motion.
PURPOSE OF THE STUDY Decompression surgery with or without fusion is the gold standard treatment of lumbar spinal stenosis, but adjacent segment degeneration has been reported as a long-term complication after fusion. This led to the development of dynamic implants like the interspinous devices. They are supposed to limit extension and expand the spinal canal at the symptomatic level, but with ...
متن کاملA Finite Element Study of Mechanical Behaviour of Instrumented Lumbar Spine with Three Different Longitudinal Rods Designs: Sliding Articulated, Flexible and Rigid
Spinal fusion is the gold standard treatment for severe pathologies in lumbar spine. However, fusion can yield adverse effects, such as Adjacent Segment Degeneration (ASD). Alternatively, non-fusion techniques were proposed. Although infrequent, mechanical complications were reported such as screw loosening and screw breakages. This work aims to use Finite Element Model (FEM) to investigate imp...
متن کاملEvaluation of adjacent segments mobility after spinal instrumentation : an in vitro biomechanical study
I In nt tr ro od du uc ct ti io on n: : Spinal fusion with instrumentation is an essential component in the management of degenerative condition in the spinal motion segment. It is performed to improve initial stability with simplified external fixation for early ambulation1)2). The solid fixation, however, often induces an adjacent segment degeneration (ASD) above or below the fusion 3). ASD c...
متن کاملComparison of the Load-sharing Characteristics between Pedicle-based Composite and Rigid Rod Devices
1. Introduction Adjacent segment degeneration (ASD) has been reported as the one of late serious complication after spinal fusion surgery [1-3]. Presumably, the immediate rigidity produced by instrumentation causes more stress leading to accelerated degeneration at adjacent levels [2]. Many previous studies have been corroborated a trend toward earlier development of ASD with instrumentation as...
متن کاملDoes the Addition of a Dynamic Pedicle Screw to a Fusion Segment Prevent Adjacent Segment Pathology in the Lumbar Spine?
Study Design Retrospective clinical cohort study. Purpose To investigate whether the combined use of dynamic pedicle screws and polyaxial pedicle screws was effective on adjacent segment pathology (ASP). Overview of Literature Various screw and rod models have been recently developed for preventing adjacent segment disease, and hybrid systems have been described along with posterior instrum...
متن کامل