Assessment of scatter for the micro-CT subsystem of the trimodality FLEX Triumph preclinical scanner.
نویسندگان
چکیده
PURPOSE This work aims at assessing, through experimental measurements and Monte Carlo calculations, the scatter to primary ratio (SPR) for the micro-CT subsystem of the FLEX Triumph preclinical PET-CT scanner to improve its quantitative capabilities. METHODS Experimental measurements were carried out using the single blocker method, where five cylindrical blockers with diameters ranging between 3 and 11.65 mm were used to assess the SPR without the blocker through interpolation. Because of the vertical layout of the imaging device, the blocker was placed over rat-sized and mouse-sized phantoms and central and peripheral SPR values were obtained by rotating the source and detector. The influence of beam energy (30, 50, and 80 kVp), geometrical magnification (1.3 and 2.0) and phantom diameter (25 and 50 mm) and density (polyethylene and water) were investigated. Monte Carlo (MC) simulations using the MCNP4C code were also performed and compared to experimental results to validate their accuracy. RESULTS The highest difference was found in the extreme peripheral region of the small phantom, while the maximum difference at the center of the phantom is about 6%, indicating that MC simulations can reproduce well the experimental results, at least in the region inside the phantom. The maximal SPR (0.562) was obtained for the large phantom at 30 kVp and a magnification of 1.3. The full SPR profile was calculated using MC simulations and used to express its dependency on beam energy (quadratic), air gap (asymptotic), and phantom diameter (quadratic). CONCLUSIONS The obtained results are in good agreement with theoretical predictions. MC simulations were valuable for the evaluation of the influence of various acquisition parameters on the SPR estimates.
منابع مشابه
Performance comparison of four commercial GE discovery PET/CT scanners: A monte carlo study using GATE
Combined PET/CT scanners now play a major role in medicine for in vivo imaging in oncology, cardiology, neurology, and psychiatry. As the performance of a scanner depends not only on the scintillating material but also on the scanner design, with regards to the advent of newer scanners, there is a need to optimize acquisition protocols as well as to compare scanner ...
متن کاملNEMA NU-04-based Performance Characteristics of the LabPET-8TM Small Animal PET Scanner
The objective of this study is to characterize the performance of the preclinical avalanche photodiode (APD)based LabPET-8TM subsystem of the fully integrated trimodality PET/SPECT/CT TriumphTM scanner using the NEMA NU 04 – 2008 protocol. The characterized performance parameters include the spatial resolution, sensitivity, scatter fraction, counts rate performance, and image quality characteri...
متن کاملCalculation of the Scattered Radiation Profile in 64 Slice CT Scanners Using Experimental Measurement
Introduction: One of the most important parameters in x-ray CT imaging is the noise induced by detected scattered radiation. The detected scattered radiation is completely dependent on the scanner geometry as well as size, shape and material of the scanned object. The magnitude and spatial distribution of the scattered radiation in x-ray CT should be quantified for development of robust scatter...
متن کاملNEMA NU-04-based performance characteristics of the LabPET-8™ small animal PET scanner.
The objective of this study is to characterize the performance of the preclinical avalanche photodiode (APD)-based LabPET-8™ subsystem of the fully integrated trimodality PET/SPECT/CT Triumph™ scanner using the National Electrical Manufacturers Association (NEMA) NU 04-2008 protocol. The characterized performance parameters include the spatial resolution, sensitivity, scatter fraction, counts r...
متن کاملEffect of scatter coincidences, partial volume, positron range and non-colinearity on the quantification of FDOPA Patlak analysis
Introduction: The key characteristics of positron emission tomography (PET) are its quantitative capability and its sensitivity, which allow the in vivo imaging of biochemical interactions with small amounts of tracer concentrations. Therefore, accurate quantification is important. However, it can be sensitive to several physical factors. The aim of this investigation is the assessment of the e...