Sensitivity of array detector measurements in determining shifts of MLC leaf positions

نویسندگان

  • Qingyang Shang
  • Andrew Godley
  • Long Huang
  • Peng Qi
  • Ping Xia
چکیده

Using a MatriXX 2D ionization chamber array, we evaluated the detection sensitivity of systematically introduced MLC leaf positioning shifts to test whether the conventional IMRT QA method can be used for quality assurance of an MLC tracking algorithm. Because of finite special resolution, we first tested whether the detection sensitivity was dependent of the locations of leaf shifts and positions of ionization chambers. We then introduced the same systematic leaf shifts in two clinical intensity modulated radiotherapy plans (prostate and head and neck cancer). Our results reported differences between the measured planar doses with and without MLC shifts (errors). Independent of the locations of the leaf position shifts and positions of the detectors, for the simple rectangular fields, the MatriXX was able to detect ±2 mm MLC leaf positioning shifts with Gamma index of 3%/3 mm and ±1 mm MLC leaf position shifts with Gamma index of 2%/2 mm. For the clinical plans, measuring the fields individually, leaf positioning shifts of ±2 mm were detected using Gamma index of 3%/3 mm and a passing rate of 95%. When the fields were measured compositely, the Gamma index exhibited less sensitivity for the detection of leaf positioning shifts than when the fields were measured individually. In conclusion, if more than 2 mm MLC leaf shifts were required, the commercial detector array (MatriXX) is able to detect such MLC positioning shifts, otherwise a more sensitive quality assurance method should be used.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

On the sensitivity of patient‐specific IMRT QA to MLC positioning errors

Accurate multileaf collimator (MLC) leaf positioning plays an essential role in the effective implementation of intensity modulated radiation therapy (IMRT). This work evaluates the sensitivity of current patient-specific IMRT quality assurance (QA) procedures to minor MLC leaf positioning errors. Random errors of up to 2 mm and systematic errors of +/-1 mm and +/-2 mm in MLC leaf positions wer...

متن کامل

A method to enhance spatial resolution of a 2D ion chamber array for quality control of MLC

This work introduces a new method for verifying MLC leaf positions with enough spatial resolution to replace film-based methods in performing QA tests. It is implemented on a 2D ion chamber array, and it is based on the principle of varying signal response of a volumetric detector to partial irradiation. A PTW 2D-ARRAY seven29 (PTW-729 2D) array was used to assess a Siemens OPTIFOCUS MLC. Parti...

متن کامل

Evaluation of two methods of predicting MLC leaf positions using EPID measurements.

In intensity modulated radiation treatments (IMRT), the position of the field edges and the modulation within the beam are often achieved with a multileaf collimator (MLC). During the MLC calibration process, due to the finite accuracy of leaf position measurements, a systematic error may be introduced to leaf positions. Thereafter leaf positions of the MLC depend on the systematic error introd...

متن کامل

Maximum MLC opening effect in dynamic delivery of IMRT: leaf‐positional analysis

The analysis of dynamic multileaf collimator (MLC) positions for the delivered intensity-modulated radiotherapy (IMRT) plans is crucial in that it may capture dose delivery problems otherwise difficult to observe and quantify in the conventional dosimetric measurements with film or with an ionization chamber. In some IMRT systems, delivery of IMRT fields starts with a maximum MLC opening (rough...

متن کامل

A clinically observed discrepancy between image-based and log-based MLC positions.

PURPOSE To present a clinical case in which real-time intratreatment imaging identified an multileaf collimator (MLC) leaf to be consistently deviating from its programmed and logged position by >1 mm. METHODS An EPID-based exit-fluence dosimetry system designed to prevent gross delivery errors was used to capture cine during treatment images. The author serendipitously visually identified a ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 18  شماره 

صفحات  -

تاریخ انتشار 2017