Comparison of different inspiratory triggering settings in automated ventilators during cardiopulmonary resuscitation in a porcine model

نویسندگان

  • Dingyu Tan
  • Jun Xu
  • Shihuan Shao
  • Yangyang Fu
  • Feng Sun
  • Yazhi Zhang
  • Yingying Hu
  • Joseph Walline
  • Huadong Zhu
  • Xuezhong Yu
چکیده

BACKGROUND Mechanical ventilation via automated in-hospital ventilators is quite common during cardiopulmonary resuscitation. It is not known whether different inspiratory triggering sensitivity settings of ordinary ventilators have different effects on actual ventilation, gas exchange and hemodynamics during resuscitation. METHODS 18 pigs enrolled in this study were anaesthetized and intubated. Continuous chest compressions and mechanical ventilation (volume-controlled mode, 100% O2, respiratory rate 10/min, and tidal volumes 10ml/kg) were performed after 3 minutes of ventricular fibrillation. Group trig-4, trig-10 and trig-20 (six pigs each) were characterized by triggering sensitivities of 4, 10 and 20 (cmH2O for pressure-triggering and L/min for flow-triggering), respectively. Additionally, each pig in each group was mechanically ventilated using three types of inspiratory triggering (pressure-triggering, flow-triggering and turned-off triggering) of 5 minutes duration each, and each animal matched with one of six random assortments of the three different triggering settings. Blood gas samples, respiratory and hemodynamic parameters for each period were all collected and analyzed. RESULTS In each group, significantly lower actual respiratory rate, minute ventilation volume, mean airway pressure, arterial pH, PaO2, and higher end-tidal carbon dioxide, aortic blood pressure, coronary perfusion pressure, PaCO2 and venous oxygen saturation were observed in the ventilation periods with a turned-off triggering setting compared to those with pressure- or flow- triggering (all P<0.05), except when compared with pressure-triggering of 20 cmH2O (respiratory rate 10.5[10/11.3]/min vs 12.5[10.8/13.3]/min, P = 0.07; coronary perfusion pressure 30.3[24.5/31.6] mmHg vs 27.4[23.7/29] mmHg, P = 0.173; venous oxygen saturation 46.5[32/56.8]% vs 41.5[33.5/48.5]%, P = 0.575). CONCLUSIONS Ventilation with pressure- or flow-triggering tends to induce hyperventilation and deteriorating gas exchange and hemodynamics during CPR. A turned-off patient triggering or a pressure-triggering of 20 cmH2O is preferred for ventilation when an ordinary inpatient hospital ventilator is used during resuscitation.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Effectiveness of Inspiratory Termination Synchrony with Automatic Cycling During Noninvasive Pressure Support Ventilation

BACKGROUND Pressure support ventilation (PSV) is a standard method for non-invasive home ventilation. A bench study was designed to compare the effectiveness of patient-ventilator inspiratory termination synchronization with automated and conventional triggering in various respiratory mechanics models. MATERIAL AND METHODS Two ventilators, the Respironics V60 and Curative Flexo ST 30, connected...

متن کامل

Performance of transport ventilator with patient-triggered ventilation.

OBJECTIVES Transport ventilators with inspiratory triggering functions and pressure support-control modes have recently become commercially available. We evaluated these ventilators in comparison with a standard ICU ventilator. STUDY DESIGN Laboratory study with a mechanical lung model. METHODS We compared the performance of four transport ventilators (model 740, Mallinckrodt, Pleasanton, C...

متن کامل

Effects of inspiratory rise time on triggering work load during pressure-support ventilation: a lung model study.

BACKGROUND The rise in inspiratory flow is important during patient-triggered ventilation. Many ventilators incorporate a function to control the time to reach the targeted airway pressure (inspiratory rise time). However, it has not been clarified how inspiratory rise time affects inspiratory work load under various ventilator settings. In a bench study we investigated the effect of inspirator...

متن کامل

Effects of gas leak on triggering function, humidification, and inspiratory oxygen fraction during noninvasive positive airway pressure ventilation.

OBJECTIVES During noninvasive positive pressure ventilation (NPPV), the gas leak that commonly occurs around the mask can render NPPV ineffective. We evaluated the effects of gas leak on inspiratory trigger function during NPPV with bilevel pressure and ICU ventilators. In addition, we evaluated the effects of gas leak on fraction of inspired oxygen (Fio(2)) and humidification. METHODS Air le...

متن کامل

Performance characteristics of five new anesthesia ventilators and four intensive care ventilators in pressure-support mode: a comparative bench study.

BACKGROUND During the past few years, many manufacturers have introduced new modes of ventilation in anesthesia ventilators, especially partial-pressure modalities. The current bench test study was designed to compare triggering and pressurization of five new anesthesia ventilators with four intensive care unit ventilators. METHODS Ventilators were connected to a two-compartment lung model. O...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 12  شماره 

صفحات  -

تاریخ انتشار 2017