Plutonium, protactinium, uranium and thorium isotopes determination in environmental samples by SF ICP-MS
نویسندگان
چکیده
Alpha spectrometry is the most commonly used analytical tool for the determination of Pu, Am, Pa, U and Th long live radionuclides. Thermal ionization mass spectrometry (TIMS) and acceleration mass spectrometry (AMS) are also still considered as primary methods for the determination of their isotope ratios. Spectrometers with plasma ion sources now offer another efficient alternative method with the development of sector field inductively coupled plasma mass spectrometry (SF ICP-MS). This paper summarizes the development work performed and results obtained using this technique for the analysis of such radionuclides. Isotopic ratios and ultra trace levels of man made radionuclides like 239, 240, 241 plutonium but also naturally occurring species like 231 protactinium, 234, 235, 238 uranium and 230, 232 thorium were determined in different field samples and reference materials using a micro concentric nebulizer as an introduction system for the mass spectrometer. Prior to the mass spectrometric measurements these elements were isolated from major elements by different selective precipitations. In further steps, they were purified by anion exchange. Results of isotope ratios including plutonium ones are also discussed.
منابع مشابه
Determination of 210Po and uranium in high salinity water samples
A method for the determination of uranium and 210Po in high salinity water samples has been elaborated. Both radionuclides are preconcentrated from 0.5 dm3 saline media by co-precipitation with hydrated manganese dioxide, followed by dissolution of the precipitate in 200 mL of 1 M HCl. Uranium isotopes 235U and 238U can be directly determined by ICP MS method with a detection limit of 0.01 ppb ...
متن کاملDetermination of Actinide Elements in Environmental Samples by Icp-ms
DETERMINATION OF ACTINDE ELEMENTS iN ENVIRONMENTAL SAMPLES BY ICP-MS Jason Bedford Truscott Methods for the determination of the actinide elements m water, biological, soil and sediment samples have been developed usmg on-line sohd phase extraction and high performance liquid chromatography (I{PLC) coupled with mductively coupled plasma mass spectrometry (ICP-MS) Initial applications utilised a...
متن کاملC2ja10304d 327..334
An on-line solid-phase extraction method linked to inductively coupled mass spectrometry (ICP-MS) has been developed for the determination of low levels of uranium and thorium in environmental samples. The hyphenation of lab-on-valve (LOV) and multisyringe flow injection analysis (MSFIA), coupled to an ICP-MS, allows the simultaneous determination of thorium and uranium in different types of en...
متن کاملValidation of Bulk Analysis with Simulated Swipe Samples Containing Ultra- Trace Amounts of Uranium and Plutonium Using MC-ICP-MS
Suitable analytical procedures for the bulk analysis of ultra-trace amounts of uranium and plutonium have been developed using multi-collector inductively coupled mass spectrometry (MC–ICP–MS). The quantification and determination of the isotopic ratios of uranium and plutonium in three simulated swipe samples, a swipe blank, and a process blank were performed to validate the analytical perform...
متن کاملCritical comparison of radiometric and mass spectrometric methods for the determination of radionuclides in environmental, biological and nuclear waste samples.
The radiometric methods, alpha (alpha)-, beta (beta)-, gamma (gamma)-spectrometry, and mass spectrometric methods, inductively coupled plasma mass spectrometry, accelerator mass spectrometry, thermal ionization mass spectrometry, resonance ionization mass spectrometry, secondary ion mass spectrometry, and glow discharge mass spectrometry are reviewed for the determination of radionuclides. Thes...
متن کامل