Early failure of the Ultima metal-on-metal total hip replacement in the presence of normal plain radiographs.
نویسندگان
چکیده
Metal-on-metal total hip replacement has been targeted at younger patients with anticipated long-term survival, but the effect of the production of metal ions is a concern because of their possible toxicity to cells. We have reviewed the results of the use of the Ultima hybrid metal-on-metal total hip replacement, with a cemented polished tapered femoral component with a 28 mm diameter and a cobalt-chrome (CoCr) modular head, articulating with a 28 mm CoCr acetabular bearing surface secured in a titanium alloy uncemented shell. Between 1997 and 2004, 545 patients with 652 affected hips underwent replacement using this system. Up to 31 January 2008, 90 (13.8%) hips in 82 patients had been revised. Pain was the sole reason for revision in 44 hips (48.9%) of which 35 had normal plain radiographs. Peri-prosthetic fractures occurred in 17 hips (18.9%) with early dislocation in three (3.3%) and late dislocation in 16 (17.8%). Infection was found in nine hips (10.0%). At operation, a range of changes was noted including cavities containing cloudy fluid under pressure, necrotic soft tissues with avulsed tendons and denuded osteonecrotic upper femora. Corrosion was frequently observed on the retrieved cemented part of the femoral component. Typically, the peri-operative findings confirmed those found on pre-operative metal artefact reduction sequence MRI and histological examination showed severe necrosis. Metal artefact reduction sequence MRI proved to be useful when investigating these patients with pain in the absence of adverse plain radiological features.
منابع مشابه
Early failure of articular surface replacement XL total hip arthroplasty.
The ASR (articular surface replacement) XL (DePuy, Warsaw, Ind) metal-on-metal hip arthroplasty offers the advantage of stability and increased motion. However, an alarming number of early failures prompted the evaluation of patients treated with this system. A prospective study of patients who underwent arthroplasty with the ASR XL system was performed. Patients with 2-year follow-up or any re...
متن کاملTotal Knee Arthroplasty Failure Induced by Metal Hypersensitivity
BACKGROUND Metal hypersensitivity is an uncommon complication after total knee arthroplasty (TKA) that can lead to significant functional impairment and aseptic prosthesis failure. CASE REPORT We describe a 70-year-old patient who presented with persistent pain, swelling, and instability 2 years after a primary TKA. The patient had a history of metal hypersensitivity following bilateral metal...
متن کاملMetal in Total Hip Arthroplasty: Wear Particles, Biology, and Diagnosis.
Total hip arthroplasty (THA) has been performed for nearly 50 years. Between 2006 and 2012, more than 600,000 metal-on-metal THA procedures were performed in the United States. This article reviews the production of metal wear debris in a metal-on-metal articulation and the interaction of cobalt and chromium ions that ultimately led to a dramatic decline in the use of metal-on-metal THA articul...
متن کاملA comparison of leg length and femoral offset discrepancies in hip resurfacing, large head metal-on- metal and conventional total hip replacement: a case series
BACKGROUND A discrepancy in leg length and femoral offset restoration is the leading cause of patient dissatisfaction in hip replacement surgery and has profound implications on patient quality of life. The aim of this study is to compare biomechanical hip reconstruction in hip resurfacing, large-diameter femoral head hip arthroplasty and conventional total hip replacement. METHOD Sixty patie...
متن کاملTotal Hip Replacement Revision in a Single Brand Small Cementless Stem – Our Experience after the Findings of the National Joint Registry
Background: Cementless total hip replacement is the common THR performed in England, Wales, Northern Irelandand the Isle of Man. The Corail stem is the most popular cementless implant and has a ODEP 10A rating. Review ofits performance in the registry identified an increase rate of revision amongst the smaller stem sizes. However, claritywas not provided on the explanation for this finding. We ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- The Journal of bone and joint surgery. British volume
دوره 92 11 شماره
صفحات -
تاریخ انتشار 2010