In vivo assessment of the impedance ratio method used in electronic foramen locators
نویسندگان
چکیده
BACKGROUND The results of an in vivo study on the "ratio method" used in electronic foramen locators (EFL) are presented. EFLs are becoming widely used in the determination of the working length (WL) during the root canal treatment. The WL is the distance from a coronal reference point to the point at which canal preparation and filling should terminate. The "ratio method" was assessed by many clinicians with the aim of determining its ability to locate the apical foramen (AF). Nevertheless, in vivo studies to assess the method itself and to explain why the "ratio method" is able to locate the apical foramen and is unable to determine intermediate distances were not published so far. METHODS A developed apparatus applies an electrical current signal with constant amplitude of 10 μARMS through the endodontic file within the root canal. The applied current signal is composed by summing six sine waves, from 250 Hz to 8 kHz. Data were acquired with the endodontic file tip at 7 different positions within root canals. In the frequency domain the quotients between the amplitude of a reference frequency and the amplitudes of the other frequencies components were calculated. Twenty one root canals were analyzed in vivo, during the endodontic treatment of twelve teeth of different patients, with age between 20 to 55 years. RESULTS For the range of frequencies used in the commercial EFLs and for distances ranging from -3 mm to -1 mm of the AF, the impedance of the root canal is mainly resistive. However, when the file tip gets closer to AF, the root canal electrical impedance starts to change from a mainly resistive to a complex impedance. This change in the measured root canal impedance starts when the file tip is near -1.0 mm from the AF, getting stronger as the file tip gets closer to the AF. This change in the impedance behavior affects the ratio (quotient) of the impedance measured at different frequencies. Through graphic analysis it is demonstrated why EFLs based on the ratio method are unable to accurately measure any distances between - 3.0 and -0.5 mm from the apical foramen. The only reliable measurement is the 0 mm distance, which is when the file tip is at the AF. CONCLUSIONS The electrical impedance values of 21 root canals were in vivo studied. The results confirm the ability of EFLs that are based on the ratio method to accurately locate the AF position and explain why they are unable to determine the file tip position along the root canal.
منابع مشابه
In vitro evaluation of the accuracy of two new electronic apex locators in root canal length determination
In vitro evaluation of the accuracy of two new electronic apex locators in root canal length determination Dr. Hasheminia, S.M.* - Dr. Jafari, N.** *Associate Professor of Endodontics Dept., Faculty of Dentistry, Isfahan University of Medical Sciences. ** Dentist. Abstract Background & Aim: Radiography is still the common method of measuring working length in root canal therapy. While, gag refl...
متن کاملAn in vivo comparison of gradient and absolute impedance electronic apex locators.
Two electronic apex locators based on the gradient (Apit) and absolute (Odontometer) impedance principles were evaluated. Distances of file tips to apical foramina were determined in vivo with the electronic apex locators and subsequently verified after extraction. Mean differences between file tips and apical foramina were +0.14 +/- 0.27 mm and -0.36 +/- 0.71 mm for the Apit and Odontometer, r...
متن کاملThe effect of different needle-insertion depths on the accuracy of the Root ZX II and Root ZX mini apex locators in the presence of various irrigants
BACKGROUND AND AIM: The aim of the present study was to evaluate the effect of different needle-insertion depths on the accuracy of the Root ZX II and Root ZX mini apex locators in the presence of various irrigants. METHODS: Ninety extracted single-canal human teeth were used in nine experimental groups. The coronal one-third of the canals was preflared, and the length to the major foramen was ...
متن کاملEx vivo comparison of two electronic apex locators with different scales and frequencies.
AIM To compare ex vivo the accuracy of two impedance quotient apex locators with different scales and frequencies of the measuring circuit. METHODOLOGY In each root of 193 extracted human teeth, electronic working length determination (ELD) was carried out with a newly constructed measuring unit. In all cases, ELD was performed using the apex locators Justy II (Hager & Werken, Duisburg, Germa...
متن کاملارزیابی دقت یک نوع Electronic Apex Locator در تعیین طول کانال دندانهای شیری در شرایط In vivo
Background and Aim: Radiography is the most commonly used technique in root canal length determination, but its application in pedodontics is difficult due to many problems such as radiation hazards, superimposition of permanent teeth buds and uncooperative children. The aim of this study was to investigate the accuracy of an electronic apex locator (Dentaport ZX) in measuring the root canal le...
متن کامل