Investigation of the clinical performance of a novel solid‐state diagnostic dosimeter
نویسندگان
چکیده
This study investigated the clinical performance of a novel solid-state diagnostic dosimeter, the RaySafe Xi transparent detector, by comparing its performance to a reference-class ionization chamber. Firstly a comparison of dosimeter response "free-in-air" with standard beam qualities was made, followed by an investigation into its relative transparency in an X-ray field and angular sensitivity dependence. The second part of the study looked at the overall performance of the transparent detector under scatter conditions with a number of beam qualities, including standard beam and those hardened by copper (Cu) filtration of thickness up to 0.9 mm, as would be encountered in the equipment testing of fluoroscopy systems. Overall, the transparent detector has demonstrated equivalent measurement properties to the ionization chamber under standard conditions and provided similar X-ray attenuation as reflected by the nearly identical radiographic parameters selected for both dosimeters by the automatic dose rate control (ADRC) system. Yet, it also possessed an asymmetric angular response which respectively under- and overestimated the dose contribution from the rear and lateral directions by the same amount of 50%. The transparent detector provided comparable dose reading of ± 3% to the ionization chamber with standard beam qualities and backscatter radiation present. These results were in good agreement with those of free-in-air measurement, indicating that the angular under- and overresponse might potentially compensate one another for accurate measurement. However, for identical Cu filtered beam qualities and setups, the transparent detector on average overresponded by 5.4% across the useful tube voltage range. In conclusion, the transparent detector, with its novel design, is essentially equivalent, within a 5% tolerance, to an ionization chamber, except in situations where beams hardened with Cu filtration are used with backscatter radiation present requiring larger uncertainty error estimations.
منابع مشابه
Experimental evaluation of thermal stability of PS- MWCNT nanocomposite as a real-time dosimeter
Introduction: Carbon nanotubes (CNTs) are ideal reinforcing fillers for a polymer matrix, because of their nanometer size, high aspect ratio and, more importantly, their excellent mechanical strength, and electrical and thermal conductivity. The CNTs are categorized into two groups: SWCNT and MWCNT. MWCNTs would have diameters ranging from 2 to 100 nm and lengths of up to tens ...
متن کاملSynthesis of strontium borate doped with Pb by solid state method and characterization of its thermoluminescent glow curve
Strontium Tetraborate doped with Lead SrB4O7:Pb powder was prepared by solid state method. The formation of Strontium Tetraborate compound was analyzed by X-ray diffraction (XRD) and Scanning Electron Microscopy (SEM). Thermoluminescence Glow Curve of gamma irradiated sample was investigated in order to use as a thermoluminescent dosimeter. Because of excellent mechanical properties, non-hygros...
متن کاملA Validation Study on Radiation properties of a Novel LiF: Mg, Ti Known as IAP-100
Introduction: LiF dosimeter has the most application in medicine. This study aimed to evaluate some dosimetric properties of a novel LiF: Mg, Ti. Materials and Methods: An ELEKTA Precise linear accelerator was used to calibrate dosimeters at 6 MV. In this survey, responses of dosimeters were evaluated up to 1000 cGy. Background effect was investigated in two different dosimeter states includin...
متن کاملLevels of natural radioactivity in environment in residential area of Moradabad District, Western Uttar Pradesh
Indoor radon and thoron have been measured in the houses ofMoradabad District, Uttar Pradesh India, by means of solid state nuclear trackdetectors. Radon, an invisible radioactive gas, occurs naturally in indooratmospheres and along with thoron is the most important contribution of humanexposure to natural sources. Radon exists in soil gas building materials and indooratmosphere to name but a f...
متن کاملDiagnostic Utility of miRNAs in Cancer
Cancer is the one of most prevalent and leading causes of death in the world. Current ad­vancements in technology improve the understanding of the pathogenesis and pathology of cancers. But, due to enlarging mortality rates, poor prognosis, and lacunae in clinical early predictive biomarkers provide an important momentum to investigate novel early diagnos­tic/prognostic markers and spec...
متن کامل