Automatic polyp detection and measurement with computed tomographic colonography: A phantom study
نویسندگان
چکیده
PURPOSE The purpose of this study is to assess the performance of computer-aided detection (CAD) software in detecting and measuring polyps for CT Colonography, based on an in vitro phantom study. MATERIAL AND METHODS A colon phantom was constructed with a PVC pipe of 3.8 cm diameter. Nine simulated polyps of various sizes (3.2mm-25.4mm) were affixed inside the phantom that was placed in a water bath. The phantom was scanned on a 64-slice CT scanner with tube voltage of 120 kV and current of 205 mAs. Two separate scans were performed, with different slice thickness and reconstruction interval. The first scan (thin) had a slice thickness of 1mm and reconstruction interval 0.5mm. The second scan (thick) had a slice thickness of 2mm and reconstruction interval of 1mm. Images from both scans were processed using CT Colonography software that automatically segments the colon phantom and applies CAD that automatically highlights and provides the size (maximum and minimum diameters, volume) of each polyp. Two readers independently measured each polyp (two orthogonal diameters) using both 2D and 3D views. Readers' manual measurements (diameters) and automatic measurements from CAD (diameters and volume) were compared to actual polyp sizes as measured by mechanical calipers. RESULTS All polyps except the smallest (3.2mm) were detected by CAD. CAD achieved 100% sensitivity in detecting polyps ≥6mm. Mean errors in CAD automated volume measurements for thin and thick slice scans were 8.7% and 6.8%, respectively. Almost all CAD and manual readers' 3D measurements overestimated the size of polyps to variable extent. Both over- and underestimation of polyp sizes were observed in the readers' manual 2D measurements. Overall, Reader 1 (expert) had smaller mean error than Reader 2 (non-expert). CONCLUSION CAD provided accurate size measurements for all polyps, and results were comparable to the two readers' manual measurements.
منابع مشابه
Segmentation and Size Measurement of Polyps in CT Colonography
Virtual colonoscopy is a relatively new method for the detection of colonic polyps. Their size, which is measured from reformatted CT images, mainly determines diagnosis. We present an automatic method for measuring the polyp size. The method is based on a robust segmentation method that grows a surface patch over the entire polyp surface starting from a seed. Projection of the patch points alo...
متن کاملDevelopment of a synthetic phantom for the selection of optimal scanning parameters in CAD-CT colonography.
The aim of this paper is to present the development of a synthetic phantom that can be used for the selection of optimal scanning parameters in computed tomography (CT) colonography. In this paper we attempt to evaluate the influence of the main scanning parameters including slice thickness, reconstruction interval, field of view, table speed and radiation dose on the overall performance of a c...
متن کاملA Complete System for Polyps Flagging in Virtual Colonoscopy
Computer tomographic colonography, combined with computer-aided detection, is a promising emerging technique for colonic polyp analysis. We present a complete pipeline for polyp flagging based on a simple segmentation technique that enhances polyps, a multi-scale candidate polyp delineation, and new texture and geometric features that consider both the information in the candidate polyp locatio...
متن کاملPerformance of computed tomographic colonography for the screening of colorectal polyp in acromegalic patients: a prospective study.
CONTEXT Acromegalic patients have better chances to develop colorectal polyps and cancer and, considered a high-risk group, need to undergo frequent screening examinations. Moreover, in acromegalia, the increased bowel length and the intestinal loop complexity can lead to higher levels of technical difficulties and increase the risks of complications at conventional colonoscopy. Computed tomogr...
متن کاملComputer-aided diagnosis scheme for detection of polyps at CT colonography.
Colon cancer is one of the leading causes of cancer deaths in the United States. However, most colon cancers can be prevented if precursor colonic polyps are detected and removed. An advanced computer-aided diagnosis (CAD) scheme was developed for the automated detection of polyps at computed tomographic (CT) colonography. A region encompassing the colonic wall is extracted from an isotropic vo...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 5 شماره
صفحات -
تاریخ انتشار 2009