Comparison of Architect I 2000 for Determination of Cyclosporine with Axsym
نویسندگان
چکیده
BACKGROUND Cyclosporine has been shown effective drug in suppressing acute rejection in recipients of allograft organ transplants. METHODS The cyclosporine concentration of 96 blood samples was determined using CMIA (chemiluminesecent microparticle immnoassay) Architect i 2000 and FPIA (fluorescence polarization immunoassay) AxSYM Abbott diagnostic. All patients have transplantation of kidneys and were hospitalized at Department of Nephrology at the Clinical center of University of Sarajevo. The reference serum range of cyclosporine for kidney organ transplantation for maintenance lies between 50 and 150 ng/mL. The quality control, precision and accurancy of Architect i 2000 were assessed. RESULTS The quality control was done using quality control serums for low (= 91 ng/mL), medium (= 328 ng/mL) and high (= 829 ng/mL). We have used commercial BIORAD controls and got reproducibility CV 5.83 % to 13 % for Architect i 2000. It was established that the main difference between Architect i 2000 and AxSYM and it was statistically significant for P < 0.05 according to Student t-test. Correlation coefficient was r = 0.903. CONCLUSION The CMIA Architect assay has significant reduced cyclosporine metabolite interference relative to other immunoassay and is a convenient and sensitive automated method to measure cyclosporine in whole blood.
منابع مشابه
Comparison of ARCHITECT chemiluminiscent microparticle immunoassay for determination of Troponin I in serum with AXYM MEIA technology
Introduction: The aim of this study was determination of troponin I at serum using Architect (Abbott) and AxSYM System (Abbott). Troponin is regulatory subunit of the troponin complex associate with actin filament within muscle cells and it is a marker for diagnosis of myocardial damage. Methods: We used Architect STAT chemiluminescent microparticle immunoassay (CMIA) and AxSYM microparticle En...
متن کاملPerformance characteristics of the ARCHITECT Active-B12 (Holotranscobalamin) assay.
BACKGROUND Vitamin B12 (cobalamin) is a necessary cofactor in methionine and succinyl-CoA metabolism. Studies estimate the deficiency prevalence as high as 30% in the elderly population. Ten to thirty percent of circulating cobalamin is bound to transcobalamin (holotranscobalamin, holoTC) which can readily enter cells and is therefore considered the bioactive form. The objective of our study wa...
متن کاملMultisite analytical evaluation of the Abbott ARCHITECT cyclosporine assay.
The objective of this study was to evaluate the analytical performance of the Abbott ARCHITECT Cyclosporine (CsA) immunoassay in 7 clinical laboratories in comparison to liquid chromatography/tandem mass spectrometry (LC/MS/MS), Abbott TDx, Cobas Integra 800, and the Dade Dimension Xpand immunoassay. The ARCHITECT assay uses a whole blood specimen, a pretreatment step with organic reagents to p...
متن کاملComparing the digoxin test with the Architect® i1000sr system with respect to the AxSYM® system.
OBJECTIVE To assess the technique employed by the autoanalyser Architect® i1000sr to determine digoxin in serum samples, compared with the assay developed for AsSYM® using microparticle enzyme immunoassay (DigoxinII). METHOD A prospective analysis of the samples from 100 requests to monitor patients being treated with digoxin. The samples were processed in AxSYM® and Architect®. The technique...
متن کاملLower cardiac troponin T and I results in heparin-plasma than in serum.
BACKGROUND The use of plasma rather than serum for determination of cardiac troponins can improve turnaround time and potentially avoid incomplete serum separation that may produce falsely increased results. We investigated the influence of incomplete serum separation and the effect of heparin-plasma on cardiac troponin concentrations. METHODS Serum and heparin-plasma samples were drawn simul...
متن کامل