BACKGROUND A total of 30% to 40% of patients with congestive heart failure eligible for cardiac resynchronization therapy (CRT) either do not respond to conventional CRT or remain untreated due to an inability or impediment to coronary sinus (CS) lead implantation. The WiSE-CRT system (EBR Systems, Sunnyvale, California) was developed to address this at-risk patient population by performing biventricular pacing via a wireless left ventricular (LV) endocardial pacing electrode. OBJECTIVES The SELECT-LV (Safety and Performance of Electrodes implanted in the Left Ventricle) study is a prospective multicenter non-randomized trial assessing the safety and performance of the WiSE-CRT system. METHODS A total of 35 patients indicated for CRT who had "failed" conventional CRT underwent implantation of an LV endocardial pacing electrode and a subcutaneous pulse generator. System performance, clinical efficacy, and safety events were assessed out to 6 months post-implant. RESULTS The procedure was successful in 97.1% (n = 34) of attempted implants. The most common indications for endocardial LV pacing were difficult CS anatomy (n =12), failure to respond to conventional CRT (n = 10), and a high CS pacing threshold or phrenic nerve capture (n = 5). The primary performance endpoint, biventricular pacing on the 12-lead electrocardiogram at 1 month, was achieved in 33 of 34 patients. A total of 28 patients (84.8%) had improvement in the clinical composite score at 6 months, and 21 (66%) demonstrated a positive echocardiographic CRT response (≥5% absolute increase in LV ejection fraction). There were no pericardial effusions, but serious procedure/device-related events occurred in 3 patients (8.6%) within 24 h, and 8 patients (22.9%) between 24 h and 1 month. CONCLUSIONS The SELECT-LV study demonstrates the clinical feasibility for the WiSE-CRT system, and provided clinical benefits to a majority of patients within an otherwise "failed" CRT population. (Safety and Performance of Electrodes Implanted in the Left Ventricle [SELECT-LV]; NCT01905670).

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Left ventricular endocardial pacing improves resynchronization therapy in canine left bundle-branch hearts.

BACKGROUND We investigated the benefits of the more physiological activation achieved by left ventricular (LV) endocardial pacing (ENDO) as compared with conventional epicardial (EPI) LV pacing in cardiac resynchronization therapy. METHODS AND RESULTS In 8 anesthetized dogs with experimental left bundle-branch block, pacing leads were positioned in the right atrium, right ventricle, and at 8 ...

متن کامل

Left ventricular endocardial or triventricular pacing to optimize cardiac resynchronization therapy in a chronic canine model of ischemic heart failure.

Cardiac resynchronization therapy (CRT) is a proven treatment for heart failure but ~30% of patients appear to not benefit from the therapy. Left ventricular (LV) endocardial and multisite epicardial [triventricular (TriV)] pacing have been proposed as alternatives to traditional LV transvenous epicardial pacing, but no study has directly compared the hemodynamic effects of these approaches. Le...

متن کامل

Development of a technique for left ventricular endocardial pacing via puncture of the interventricular septum.

BACKGROUND Left ventricular (LV) pacing through the coronary sinus is the standard approach for cardiac resynchronization therapy. When this route is unavailable, the alternatives have major limitations. LV endocardial pacing through the interventriuclar septum may offer a simpler solution. We describe an initial case series to demonstrate technical feasibility and to describe our refinement of...

متن کامل

Optimal left ventricular endocardial pacing sites for cardiac resynchronization therapy in patients with ischemic cardiomyopathy.

OBJECTIVES We sought to investigate the impact of left ventricular (LV) pacing site on mechanical response to cardiac resynchronization therapy (CRT) in patients with ischemic cardiomyopathy (ICM). BACKGROUND CRT reduces morbidity and mortality in patients with dyssynchronous LV failure; however, variability in response, particularly in ICM patients, poses ongoing challenges. Endocardial bive...

متن کامل

Left Ventricular Endocardial or Triventricular Pacing to Optimize Cardiac 1 Resynchronization Therapy in a Chronic Canine Model of Ischemic Heart

20 Background: Cardiac resynchronization therapy (CRT) is a proven treatment for heart failure 21 but approximately 30% of patients appear to not benefit from the therapy. Left ventricular (LV) 22 endocardial and multisite epicardial (TriV) pacing have been proposed as alternatives to 23 traditional LV transvenous epicardial pacing, but no study has directly compared the 24 hemodynamic effects ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Journal of the American College of Cardiology

دوره 69 17  شماره 

صفحات  -

تاریخ انتشار 2017