Fixation of intertrochanteric femur fractures using Proximal Femoral Nail Antirotation (PFNA) in the lateral decubitus position without a traction table.
نویسندگان
چکیده
OBJECTIVE The aim of this study was to evaluate whether intertrochanteric femur fractures can be reduced and nailed properly in the lateral decubitus position using Proximal Femoral Nail Antirotation (PFNA) as a fixation device without the use of a traction table. METHODS The study included 207 patients (81 male and 126 female; mean age: 75 years, range: 22 to 95 years). According to the Evans classification, there were 7 Type 1, 40 Type 2, 33 Type 3, 38 Type 4, 61 Type 5 and 28 reverse oblique fractures. Radiographs were used to measure the tip-apex distance (TAD), the quadrant of the helical blade according to Cleveland and Bosworth, Ikuta's reduction subgroup, collodiaphyseal angle and reduction gaps postoperatively. RESULTS Mean follow-up time was 20.4 (range: 6 to 38) months. According to Ikuta's classification, 176 (85%) reduced fractures were of subtype N, 15 (7.2%) subtype P and 16 (7.7%) subtype A. Good or acceptable reduction according to the Herman criteria was obtained in 99% of fractures. Mean TAD was 29.2 millimeters. Mean operation time was 57.2 minutes. Optimal blade position (center-center or inferior-center) was achieved in 53.5% of patients and was in the superior-posterior quadrants in only 2.4% of patients. Cut-out complication occurred in 9 patients (4.3%). CONCLUSION While the nailing of intertrochanteric fractures in a lateral decubitus position does not provide ideal quadrant placement and TAD, results are encouraging probably due to the excellent stability that is provided by PFNA.
منابع مشابه
Radiographic Outcomes of Osteosynthesis Using Proximal Femoral Nail Antirotation (PFNA) System in Intertrochanteric Femoral Fracture: Has PFNA II Solved All the Problems?
PURPOSE We evaluated the geometric discrepancies between the proximal femur in Koreans and two types of proximal femoral nail using plain radiographs. MATERIALS AND METHODS A total of 100 consecutive patients (38 treated with proximal femoral nail antirotation [PFNA], 62 PFNA II) with intertrochanteric fracture were retrospectively identified. The minimum follow up period was 32 months. The g...
متن کاملRandomized Controlled Trials of Proximal Femoral Nail Antirotation in Lateral Decubitus and Supine Position on Treatment of Intertrochanteric Fractures
The objective of this study was to compare the clinical results and complications of proximal femoral nail antirotation (PFNA) on treatment of intertrochanteric fractures in 120 elderly Chinese patients using Randomized Controlled Trials (RCTs). Totaly 120 cases enrolled were randomly assigned to a lateral decubitus position group and supine position group. The hospital stay, operating time, in...
متن کاملBiomechanical study on different lengths of PFNA fixation for unstable intertrochanteric femoral fractures
OBJECTIVE The aim of our study was to compare the biomechanical stability of the Proximal Femoral Nail Antirotation (PFNA) (with 200 mm, 240 mm and 280 mm-long main nails) for the management of unstable intertrochanteric femoral fractures. METHODS Tronzo-Evans Type IV and V fractures were built by applying a three-dimensional finite element model. Further, PFNA-II with 200 mm, 240 mm and 280 ...
متن کاملThe Asia proximal femoral nail antirotation versus the standard proximal femoral antirotation nail for unstable intertrochanteric fractures in elderly Chinese patients.
BACKGROUND The best options of internal fixation for unstable intertrochanteric fractures in elderly Chinese patients remain controversial. The Asia proximal femoral nail antirotation (PFNA-II) was specifically designed for Asian patients, which could be more effective than the regular proximal femoral nail antirotation (PFNA). Compared to PFNA, whether PFNA-II is associated with shorter operat...
متن کاملProximal femoral nail antirotation and third-generation Gamma nail: which is a better device for the treatment of intertrochanteric fractures?
INTRODUCTION Proximal femoral nail antirotation (PFNA) and third-generation Gamma nail (Gamma 3) are widely used in the treatment of intertrochanteric fractures. However, it remains unclear which device achieves better clinical and radiographic outcomes when treating intertrochanteric fractures. METHODS This study comprised 239 patients with intertrochanteric fractures treated with either PFN...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Acta orthopaedica et traumatologica turcica
دوره 48 5 شماره
صفحات -
تاریخ انتشار 2014