PURPOSE The goal of this study is to increase patient safety in parallel transmission (pTx) MRI systems. A major concern in these systems is radiofrequency-induced tissue heating, which can be avoided by specific absorption rate (SAR) prediction and SAR monitoring before and during the scan. METHODS In this novel comprehensive safety concept, the SAR is predicted prior to the scan based on precalculated fields obtained from electromagnetic simulations on different body models. The radiofrequency fields and the global and local SAR are supervised in real time during the scan. This concept is integrated into a 3 T pTx MR scanner and validated experimentally. RESULTS Phantom and in vivo experiments successfully validated the basic feasibility of the real-time SAR supervision concept. Supervising the SAR minimizes SAR overestimation. Monitoring the radiofrequency fields allows the detection of unsafe radiofrequency situations for the patient, which a SAR supervision system alone cannot detect. CONCLUSION This study demonstrates safe scanning in a pTx system. This new safety concept is also applicable for field strengths above 3 T and represents an important step toward safe operation of pTx systems.
منابع مشابه
Novel optically tunable capacitor for RF safe tuning and decoupling of intravascular MR coils
S. Weiss, T. Uelzen, G. Pantel, J. Mueller Philips Research Laboratories, Hamburg, Germany, Department of Micro Systems Technology, Hamburg University of Technology, Hamburg, Germany, MRI Devices Daum GmbH, Schwerin, Germany Objective: The motivation for the application of intravascular imaging coils is to gain high SNR images from a local area. To maximize SNR, proper tuning and matching of su...
متن کاملMR Physics for Physicists / B1 Shimming and Parallel Transmission
METHODS: Currently, among other means, modifications of the applied RF pulses are investigated to cope with the mentioned B1 inhomogeneities. First, using a coil array for RF transmission, standard slice selective pulses can be transmitted with different weights in the different array elements (see, e.g., [2-4]). These weights, i.e., amplitude and phase of each element, are designed to yield a ...
متن کاملImpact of 3.0 T Cardiac MR Imaging Using Dual-Source Parallel Radiofrequency Transmission with Patient-Adaptive B1 Shimming
OBJECTIVES To prospectively evaluate the impact of 3.0 T Cardiac MR imaging using dual-source parallel radiofrequency (RF) transmission with patient-adaptive B1 shimming compared with single-source RF transmission in the RF homogeneity, image contrast and image quality. METHODS The study was approved by the local institutional review board, and all subjects provided written informed consent. ...
متن کاملClinical liver MRI at 3.0 Tesla using parallel RF transmission with patient-adaptive B1 shimming
Introduction: The clinical implementation of high field MRI systems has introduced new challenges for body imaging with respect to B1 field nonuniformities, which can result in the generation of standing waves and signal loss particularly in abdominal imaging. Several approaches have been proposed to achieve a more homogeneous excitation and to avoid a degradation of image quality at higher mag...
متن کاملSAR Reduction in Parallel Transmission by k-Space Dependent RF Pulse Selection
Parallel transmission has gained considerable interest during the last few years [1,2]. The use of multiple individual RF transmit coils have been introduced to overcome B1 homogeneity limitations and to improve multi-dimensional RF pulses by shortening their duration. Of special concern in all these transmit applications is the specific absorption rate (SAR), which has to be kept below certain...