Commissioning experience with cone‐beam computed tomography for image‐guided radiation therapy
نویسندگان
چکیده
This paper reports on the commissioning of an Elekta cone-beam computed tomography (CT) system at one of the first U.S. sites to install a "regular," off-the-shelf Elekta Synergy (Elekta, Stockholm, Sweden) accelerator system. We present the quality assurance (QA) procedure as a guide for other users. The commissioning had six elements: (1) system safety, (2) geometric accuracy (agreement of megavoltage and kilovoltage beam isocenters), (3) image quality, (4) registration and correction accuracy, (5) dose to patient and dosimetric stability, and (6) QA procedures. The system passed the safety tests, and agreement of the isocenters was found to be within 1 mm. Using a precisely moved skull phantom, the reconstruction and alignment algorithm was found to be accurate within 1 mm and 1 degree in each dimension. Of 12 measurement points spanning a 9x9x15-cm volume in a Rando phantom (The Phantom Laboratory, Salem, NY), the average agreement in the x, y, and z coordinates was 0.10 mm, -0.12 mm, and 0.22 mm [standard deviations (SDs): 0.21 mm, 0.55 mm, 0.21 mm; largest deviations: 0.6 mm, 1.0 mm, 0.5 mm] respectively. The larger deviation for the y component can be partly attributed to the CT slice thickness of 1 mm in that direction. Dose to the patient depends on the machine settings and patient geometry. To monitor dose consistency, air kerma (output) and half-value layer (beam quality) are measured for a typical clinical setting. Air kerma was 6.3 cGy (120 kVp, 40 mA, 40 ms per frame, 360-degree scan, S20 field of view); half value layer was 7.1 mm aluminum (120 kV, 40 mA). We suggest performing items 1, 2, and 3 monthly, and 4 and 5 annually. In addition, we devised a daily QA procedure to verify agreement of the megavoltage and kilovoltage isocenters using a simple phantom containing three small steel balls. The frequency of all checks will be reevaluated based on data collected during about 1 year.
منابع مشابه
The Impact of Residual Geometric Inaccuracies on Normal Organ Doses in Image Guided-Radiation Therapy of Prostate Cancer Using On-Board Kilovoltage Cone-Beam Computed Tomography
Introduction: The aim of this retrospective study was to evaluate the variations in delivered dose to the bladder, rectum, and femoral heads of prostate cancer patients during a course of treatment by image-guided radiation therapy (IGRT). Materials and Methods: Overall, 15 patients with prostate cancer were selected and. Each week, for each patient five consecutive cone beam computed tomograph...
متن کاملComparative Study between Electronic Portal Imaging Device (EPID) and Cone Beam Computed Tomography (CBCT) for Radiation Treatment Verifications
Introduction: Electronic Portal Imaging Device (EPID) and Cone Beam Computed Tomography (CBCT) are the preferred tools of Image Guided Radiotherapy (IGRT) and Dose Guided Radiotherapy (DGRT) which have been used for Radiotherapy treatment verifications. As a result, the number of publications dealing with these two tools for radiation treatment verification has increased consi...
متن کاملUsability assessment of cone beam computed tomography with a full-fan mode bowtie filter compared to that with a half-fan mode bowtie filter
Background: In intensity modulated radiation therapy, cone beam computed tomography (CT) has been used to evaluate patients prior to treatment. This study conducted a comparative evaluation of the image reconstruction ability of the clinically used half-fan bowtie filter and the full-fan bowtie filter. Materals and Methods: A CT simulation marker was inserted inside a human phantom, and the pel...
متن کاملArtifact reduction techniques in Cone Beam Computed Tomography (CBCT) imaging modality
Introduction: Cone beam computed tomography (CBCT) was introduced and became more common based on its low cost, fast image procedure rate and low radiation dose compared to CT. This imaging modality improved diagnostic and treatment-planning procedures by providing three-dimensional information with greatly reduced level of radiation dose compared to 2D dental imaging modalitie...
متن کاملMethods to evaluate the performance of kilovoltage cone-beam computed tomography in the three-dimensional reconstruction space
Background: Cone-beam computed tomography (CBCT) scanners for image-guided radiotherapy are in clinical use today, but there has been no consensus on uniform acceptance to verify the CBCT image quality yet. The present work proposed new methods to fully evaluate the performance of CBCT in its three-dimensional (3D) reconstruction space. Materials and Methods: Compared to the traditional methods...
متن کاملCommissioning of Kilo-voltage Cone Beam Computed Tomography for Image-guided Radiotherapy
kV-CBCT integrated with linear accelerators as a tool for IGRT, was developed to acquire online volumetric and anatomical images (Xing 2006). The successful implementation of CBCT requires a quality assurance (QA) program that not only involves robust and intuitive quality control checks of the equipment, but also thoughtful image-guidance processes based on measurement of positional uncertaint...
متن کامل