Comparison of optical coherence reflectometry and ultrasound central corneal pachymetry.
نویسندگان
چکیده
In 50 eyes of 25 patients we prospectively measured the central corneal thickness (CCT) comparing the OLCR (Optical Low Coherence Reflectometry) pachymeter with the contact ultrasound pachymeter. The OLCR system was mounted on to a Haag-Streit slit lamp. Every single measurement was the result of 5 scans. With the contact ultrasound Sonomed pachymeter we performed 5 separate measurements and calculated the mean. The correlation between the two measurements was excellent (r = 0.99). The mean standard deviation (SD) of the measurements taken with the non-contact OLCR pachymeter was significantly lower than with the contact ultrasound pachymeter, 0.49 microm and 4.71 microm respectively (p < 0.01). The variability of the CCT measurements taken with the non-contact OLCR pachymeter is significantly lower than the variability of the CCT measurements taken with the contact ultrasound pachymetry.
منابع مشابه
Comparison of optical and ultrasound central corneal pachymetry.
In 100 eyes of 50 patients central corneal thickness was measured comparing optical to ultrasound pachymetry. With optical pachymetry the values were 1 to 40 micrometer (microns) lower in 64 eyes compared to ultrasound pachymetry. With ultrasound pachymetry the values were 1 to 18 microns lower in 26 eyes compared to optical pachymetry. In 10 cases no difference in pachymetry readings was seen ...
متن کاملCentral corneal thickness measurement with the Pentacam Scheimpflug system, optical low-coherence reflectometry pachymeter, and ultrasound pachymetry.
PURPOSE To assess the intraoperator repeatability and interoperator reproducibility of central corneal thickness measurements by the Pentacam Scheimpflug imaging system (Oculus) and the optical low-coherence reflectometer (OLCR) pachymeter (Haag-Streit) and to compare them with those of ultrasound (US) pachymetry. SETTING Assaf Harofe Medical Center Ophthalmology Outpatient Clinic, Zerifin, I...
متن کاملComparison of central corneal thickness measurements using optical low-coherence reflectometry, Fourier domain optical coherence tomography, and Scheimpflug camera.
PURPOSE To compare the results of central corneal thickness (CCT) measurements obtained using optical low-coherence reflectometry (OLCR), Fourier domain optical coherence tomography (FD-OCT), and a Scheimpflug camera (SC), combined with Placido corneal topography. METHODS A total of 25 healthy subjects were enrolled in the present study, and one eye of each subject was included. A detailed op...
متن کاملCentral corneal thickness measurements with optical coherence tomography and ultrasound pachymetry in healthy subjects and in patients after photorefractive keratectomy.
PURPOSE To compare central corneal thickness measurements made using two different methods, optical coherence tomography (OCT) and ultrasound pachymetry, applied both in normal eyes and in eyes on which photorefractive keratectomy (PRK) had been performed. A second objective was to assess the intrasession variability of OCT measurements. METHODS In this prospective study, central corneal thic...
متن کاملAccuracy of Alcon WaveLight® EX500 optical pachymetry during LASIK
PURPOSE To study the accuracy and reliability of optical pachymetry using the Alcon WaveLight EX500 during laser-assisted in situ keratomileusis (LASIK). MATERIALS AND METHODS This was a retrospective chart review of 90 eyes from 45 patients who had undergone LASIK (mean age 35.2±8.2 years; 19 males, 26 females). The WaveLight FS200 femtosecond laser was programmed to cut LASIK flaps at a des...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Bulletin de la Societe belge d'ophtalmologie
دوره 292 شماره
صفحات -
تاریخ انتشار 2004