Acuros XB Algorithm vs. Anisotropic Analytical Algorithm: A Dosimetric Study Using Heterogeneous Phantom and Computed Tomography (CT) Data Sets of Esophageal Cancer Patients

نویسندگان

  • Suresh Rana
  • Kevin Rogers
  • Shyam Pokharel
  • Terry
  • Daniel Reed
  • Christopher Biggs
چکیده

Our purpose in this study was to assess the dosimetric impact of the Acuros XB algorithm (AXB), in comparison with Anisotropic Analytical Algorithm (AAA) calculations, for esophageal cancer treatment plans created with RapidArc technique. First, we performed a phantom study by comparing the percent depth dose (PDD) calculated by AXB and AAA against the measured PDD in a slab phantom containing a 2 cm air gap thickness. Second, we performed a clinical study using a computed tomography (CT) data set from 10 esophageal cancer patients. The treatment plans calculated by AXB and AAA were evaluated for planning target volume (PTV) coverage, doses to the PTV and organs at risk (OARs). Dose calculations by the AXB and AAA were done for identical beam parameters. The AXB showed better agreement (within ±0.5%) with measurements than did the AAA (−4.9% to −6.2%). In comparison to the AAA, the AXB predicted a higher maximum PTV dose (2.0%), but lower mean (1.1%) and minimum (2.5%) PTV doses as well as reduced PTV coverage (9.1%). The averaged mean doses to all OARs predicted by the AXB were lower (up to 3.6%), and the percentage of lungs volume receiving at least 20 and 5 Gy were lower by about 3.6% in the AXB plans compared to the AAA plans. The AXB is more accurate than the AAA for dose predictions when air medium is involved. The use of AXB is more likely to avoid dose overestimation or underestimation for the esophageal cancer treatment plans compared to AAA.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Dosimetric accuracy of the Acuros XB and Anisotropic analytical algorithm near interface of the different density media for the small fields of a 6- MV flattening-filter-free beam

Background: This study was conducted to assess the accuracy of dose calculation near the air-phantom interface of a heterogeneous phantom for Acuros XB (AXB) and Anisotropic Analytical Algorithm (AAA) algorithm of a           6-MV flattening-filter-free beam, compared with film measurements. Materials and Methods: A phantom included air gap was ...

متن کامل

Dosimetric effects of the acuros XB and anisotropic analytical algorithm on volumetric modulated arc therapy planning for prostate cancer using an endorectal balloon

BACKGROUND To compare the dosimetric effects of Acuros XB (AXB) and Anisotropic Analytical Algorithm (AAA) on volumetric modulated arc therapy (VMAT) planning for postoperative prostate cancer patients irradiated using an endorectal balloon (ERB). METHODS We measured central axis doses with film in a phantom containing an air cavity, and compared measurements with calculations of the AAA and ...

متن کامل

Accuracy of the Small Field Dosimetry Using the Acuros XB Dose Calculation Algorithm within and beyond Heterogeneous Media for 6 MV Photon Beams

Purpose: The dosimetric accuracy of the recently released Acuros XB advanced dose calculation algorithm (Varian Medical Systems, Palo Alto, CA) is investigated for single radiation fields incident on homogeneous and heterogeneous geometries, as well as for two arc (VMAT) cases and compared against the analytical anisotropic algorithm (AAA), the collapsed cone convolution superposition algorithm...

متن کامل

On the dosimetric impact of inhomogeneity management in the Acuros XB algorithm for breast treatment

BACKGROUND A new algorithm for photon dose calculation, Acuros XB, has been recently introduced in the Eclipse, Varian treatment planning system, allowing, similarly to the classic Monte Carlo methods, for accurate modelling of dose deposition in media. Aim of the present study was the assessment of its behaviour in clinical cases. METHODS Datasets from ten breast patients scanned under diffe...

متن کامل

Dosimetric evaluation of Acuros XB Advanced Dose Calculation algorithm in heterogeneous media

BACKGROUND A study was realised to evaluate and determine relative figures of merit of a new algorithm for photon dose calculation when applied to inhomogeneous media. METHODS The new Acuros XB algorithm implemented in the Varian Eclipse treatment planning system was compared against a Monte Carlo method (VMC++), and the Analytical Anisotropic Algorithm (AAA). The study was carried out in vir...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2013