A clampless and sutureless aortic anastomosis technique using an endograft connector for aortoiliac occlusive disease in which the aorta cannot be clamped or sewn due to calcification or scarring.

نویسندگان

  • Dimitrios Papadimitriou
  • Dieter Mayer
  • Mario Lachat
  • Felice Pecoraro
  • Thomas Frauenfelder
  • Thomas Pfammatter
  • Hideki Ueda
  • Konstantinos Donas
  • Frank J Veith
  • Zoran Rancic
چکیده

Bypass surgery in aortoiliac or aortofemoral occlusive disease can be technically demanding and hazardous due to huge calcifications and/or patient co-morbidities. We report about mid-term results of a telescoping sutureless aortic anastomosis technique using endografts as connectors to address such challenging situations. This is a single-center experience (2004-2011) in seven patients (63 ± 6 years) requiring aortoiliac (three) or aortofemoral (four) bypass surgery. In six cases, an aortic stent graft was telescoped into the infrarenal aorta and partly deployed within the aorta and partly outside the aorta. In the first case, a bifurcated stent graft was deployed and the iliac legs were prolonged extra-anatomically with surgical grafts to reach the femoral bifurcation. In the following five cases, a tapered tubular stent graft was deployed through the aortic wall, landing inside a bifurcated surgical graft that was extra-anatomically connected to the iliac or femoral arteries. In the last case, which presented a hostile abdomen and high-risk for extensive surgery, a similar technique was used, but on the iliac artery level. In that case, an iliac stent graft re-loaded 'upside down' was deployed through the left common iliac wall, landing distally inside a hand-made 10 × 10 mm bifurcated surgical graft that was extra-anatomically connected to the left external iliac artery and to the right femoral artery. The distal anastomoses on the seven cases were performed either with running sutures (ten) or with VORTEC (four). Telescoping aortic and/or iliac anastomosis was successful in all patients. There was no perioperative mortality. One patient developed postoperative hyperperfusion of the left leg and necessitated fasciotomy. During a mean follow-up of 1.8 ± 2 years (minimum: 270 days, maximum: 7.1 years), all of the grafts remained patent and there was neither stent-graft migration nor stenosis on the level of the aortic or iliofemoral connection. One patient showed disease progression and required percutaneous transluminal angioplasty on the external iliac artery during follow-up. The uneventful perioperative course in these seven patients, with a follow-up of up to six years, underscores that this new technique can be considered in patients with aortoiliac or aortofemoral occlusive disease and in whom clamping and/or anastomosis is expected to be cumbersome or impossible.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Sutureless telescoping aortic anastomotic technique for hybrid surgical treatment of aortoiliac occlusive disease.

PURPOSE To describe a new technique to assist aortoiliac stent-graft implantation in the presence of severe and extensive calcification. TECHNIQUE The use of an aortic stent-graft telescoped into the aorta and deployed partly within and partly outside the aorta is illustrated in 2 cases. In the first, the bifurcated stent-graft was deployed with the proximal end and body intraluminally within...

متن کامل

Twelve-month patency with the PAS-port proximal connector device: a single center prospective randomized trial. Invited commentary.

BACKGROUND The PAS-Port (Cardica Inc, Redwood City, CA) is an automated system that allows for the clampless anastomosis of vein grafts to the aorta. The intent of this study was to prospectively compare one year graft patency of this system with conventional hand-sewn anastomoses in a prospective, randomized trial. METHODS A total of 99 patients undergoing elective off-pump coronary bypass s...

متن کامل

Experimental laparoscopic aortobifemoral bypass for occlusive aortoiliac disease.

OBJECTIVE To describe a totally laparoscopic technique for aortobifemoral bypass to treat aortoiliac atheromatous occlusive disease. DESIGN A feasibility study. SETTING A university teaching hospital. SUBJECTS Six piglets weighing between 70 and 80 kg were submitted to a totally laparoscopic retroperitoneal aortobifemoral bypass, performed through six trocar sites, with abdominal suspensi...

متن کامل

Experimental Model for Sutureless Proximal Anastomosis by the Viabahn Open Revascularization TEChnique (VORTEC)

Introduction In the treatment of complex aneurysms, debranching is an extra-anatomical revascularization of visceral arteries followed by endograft coverage of the thoracoabdominal aorta. It eliminates the need for a thoracotomy and aortic clamping, but requires the performance of several technically demanding visceral anastomosis. In 2008, Lachat described visceral revascularization with the u...

متن کامل

Bifurcated endograft in aortoiliac type C and D lesions: long-term results.

PURPOSE To report long-term outcome when using a bifurcated aortic endograft for treatment of aortoiliac occlusive disease (AIOD) in Trans Atlantic Inter Society Consensus (TASC) classification C and D patients. MATERIALS AND METHODS Between May 2001 and May 2009, 14 patients (11 men, 3 women) with aortoiliac TASC C and D type lesions and a mean age of 59 years ± 10 (range 41-73 years) were t...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Vascular

دوره 20 5  شماره 

صفحات  -

تاریخ انتشار 2012